Background: The utility of transcatheter application of radiofrequency energy to eliminate atrioventricular nodal reentrant tachycardia (AVNRT) was investigated.

Methods And Results: Thirty-nine patients (mean age, 53 +/- 20 years; range 14-86 years) with medically refractory AVNRT underwent perinodal ablation with radiofrequency energy. A custom-designed 6F catheter with a large (3-mm-long) distal electrode and interelectrode pacing of 2 mm was used in the majority of cases. The catheter used for ablation was initially positioned across the tricuspid anulus to obtain the largest His bundle electrogram, then withdrawn to obtain the largest atrial:ventricular electrogram ratio, with a small His bundle electrogram (less than or equal to 100 microV). Each application of radiofrequency energy (350-550 kHz, 16.2 +/- 5.2 W) was stopped after 60 seconds or if PR prolongation or an impedance rise was noted. The endpoints of the procedure were persistent modification of atrioventricular nodal conduction (either first-degree atrioventricular block or impairment of ventriculoatrial conduction) and noninducibility of AVNRT before and during isoproterenol administration. Radiofrequency energy was applied a mean of 6.8 +/- 3.5 times per session. After a mean follow-up of 8 +/- 3.0 months, 32 of the 39 patients (82%) have been free of AVNRT, and did not have high grade AV block. Three patients (8%) developed complete atrioventricular block and had pacemakers implanted. Two patients had unsuccessful initial procedures, and two patients had initially successful ablations but had recurrences of tachycardia 4-6 weeks later. Elimination of AVNRT appeared to be due to effects on the retrograde fast pathway in most patients.

Conclusions: Radiofrequency ablation of the perinodal right atrium appears to be safe and effective for treatment of typical AVNRT:

Download full-text PDF

Source
http://dx.doi.org/10.1161/01.cir.83.3.827DOI Listing

Publication Analysis

Top Keywords

radiofrequency energy
20
atrioventricular nodal
12
modification atrioventricular
8
application radiofrequency
8
bundle electrogram
8
atrioventricular block
8
atrioventricular
6
radiofrequency
6
avnrt
6
energy
5

Similar Publications

New Bipolar Electrosurgical Vessel Sealing Device Provides Improved Performance and Procedural Efficiency.

Med Devices (Auckl)

January 2025

Surgical Research and Development, Surgical Operating Unit, Medtronic, Lafayette, CO, USA.

Purpose: This study compared the performance of the new LigaSure™ XP Maryland Jaw Sealer/Divider (XP Maryland) to that of LigaSure Atlas™ (Atlas) and LigaSure™ Dolphin Tip (Dolphin Tip), two early LigaSure™ (LigaSure) devices characterized by consistent and reliable clinical performance.

Methods: Ex vivo bench testing on porcine renal arteries compared burst pressures, seal times, and rates of sticking, incomplete cuts, and charring between XP Maryland and Atlas and between XP Maryland and Dolphin Tip. In vivo acute testing on a porcine model compared thermal spread, seal times, and rates of hemostasis, sticking, and incomplete cuts between XP Maryland and the two early LigaSure devices.

View Article and Find Full Text PDF

Context: Radiofrequency ablation (RFA) is used as treatment for symptomatic thyroid nodules. Factors influencing the volume reduction ratio (VRR) at 12 months are not yet fully understood.

Objective: The primary objective was evaluating the VRR at 12 months after RFA.

View Article and Find Full Text PDF

Correlated spin-wave generation and domain-wall oscillation in a topologically textured magnetic film.

Nat Mater

January 2025

Condensed Matter Physics and Materials Science Division, Brookhaven National Laboratory, Upton, NY, USA.

Spin waves, or magnons, are essential for next-generation energy-efficient spintronics and magnonics. Yet, visualizing spin-wave dynamics at nanoscale and microwave frequencies remains a formidable challenge due to the lack of spin-sensitive, time-resolved microscopy. Here we report a breakthrough in imaging dipole-exchange spin waves in a ferromagnetic film owing to the development of laser-free ultrafast Lorentz electron microscopy, which is equipped with a microwave-mediated electron pulser for high spatiotemporal resolution.

View Article and Find Full Text PDF

Capacitive-based radiofrequency (Rf) radiation at 27 MHz offers a non-invasive approach for inducing hyperthermia, making it a promising technique for thermal cancer therapy applications. To achieve focused and site-specific hyperthermia, external material is required that efficiently convert Rf radiation into localized heat. Nanomaterials capable of absorbing Rf energy and convert into heat for targeted ablation are of critical importance.

View Article and Find Full Text PDF

Background: Concurrent Micra leadless pacemaker (Medtronic, Minneapolis, Minnesota) implantation and atrioventricular node (AVN) ablation has been shown to be feasible and safe in patients with symptomatic, drug-refractory atrial fibrillation (AF). However, major complications within the 30 days after concurrent Micra implantation and AVN ablation have been reported. We evaluated the efficacy and safety of the concurrent procedure at our institution.