A threshold photoelectron-photoion coincidence spectrometer with double velocity imaging using synchrotron radiation.

Rev Sci Instrum

Hefei National Laboratory for Physical Sciences at the Microscale, Department of Chemical Physics, University of Science and Technology of China, Hefei, Anhui 230026, People's Republic of China.

Published: November 2009

AI Article Synopsis

  • A new TPEPICO imaging spectrometer is developed at the Hefei National Synchrotron Radiation Laboratory, featuring open electron and ion lenses for simultaneous velocity mapping of photoelectrons and photoions.
  • The design includes a repelling electric field instead of a traditional accelerating field, enhancing image magnification and minimizing the impact of energetic electrons.
  • Energy and mass resolution measurements show a 9 meV resolution for TPES and a mass resolving power above 900, with a practical application demonstrated through oxygen photoionization experiments resulting in precise kinetic energy distribution measurements.

Article Abstract

A novel threshold photoelectron-photoion coincidence (TPEPICO) imaging spectrometer at the U14-A beamline of the Hefei National Synchrotron Radiation Laboratory is presented. A set of open electron and ion lenses are utilized to map velocity imaging of photoelectrons and photoions simultaneously, in which a repelling electric field using an extra lens is applied to magnify images of photoelectrons instead of traditional accelerating electric field in order to suppress the contribution of energetic electrons in the threshold photoelectron spectroscopy (TPES) and the mass-selected TPEPICO spectroscopy. The typical energy resolution of TPES is measured to be 9 meV (full width at half maximum), as shown on the (2)P(1/2) ionization of argon. The measured mass resolving power for the present TPEPICO imaging spectrometer is above 900 of M/DeltaM. Subsequently as a benchmark, oxygen molecule is photoionized by monochromatic synchrotron radiation at 20.298 eV and dissociates to an oxygen atomic ion and a neutral oxygen atom, and the translation energy distribution of oxygen atomic ion is measured by the time-sliced imaging based on mass-selected TPEPICO experiment. The kinetic energy resolution of the present ion velocity imaging is better than 3% of DeltaE/E.

Download full-text PDF

Source
http://dx.doi.org/10.1063/1.3250872DOI Listing

Publication Analysis

Top Keywords

velocity imaging
12
synchrotron radiation
12
threshold photoelectron-photoion
8
photoelectron-photoion coincidence
8
tpepico imaging
8
imaging spectrometer
8
electric field
8
mass-selected tpepico
8
energy resolution
8
oxygen atomic
8