Purpose: Transesophageal echocardiography (TEE) is the gold standard in the evaluation for left atrial appendage (LAA) thrombus in patients with atrial fibrillation (AF) and is often performed prior to AF ablation. We routinely use intracardiac echocardiography (ICE) to assist in AF ablation; however, standard right atrial views do not provide adequate visualization of the LAA. As the incidence of thrombus in this population is relatively low, TEE incurs additional risk, cost, and patient discomfort. Novel views of the LAA with ICE may obviate the need for TEE in this population. We tested the hypothesis that due to their proximity, imaging the LAA from the pulmonary artery (PA) would provide equivalent sensitivity and specificity to TEE in detecting LAA thrombus in a swine model.

Methods: Five domestic swine were utilized. Baseline images of the LAA with TEE were obtained. An 8Fr ICE catheter was placed in the left main PA, and imaging of the LAA was repeated. After transseptal puncture, an admixture of 2 cm(3) blood and 1,000 IU of thrombin was injected into the LAA, and imaging of the LAA was repeated. Two blinded, independent reviewers experienced in ICE assessed the images and adjudicated both the presence of thrombus and the subjective image quality.

Results: The presence or absence of thrombus was correctly identified in all cases by both reviewers. Both reviewers rated the subjective quality of ICE images superior to TEE.

Conclusions: ICE is equivalent to TEE in imaging LAA thrombus in a porcine model. Whether ICE can provide similar diagnostic accuracy and safety for detecting LAA thrombus in humans remains unproven.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10840-009-9450-3DOI Listing

Publication Analysis

Top Keywords

laa thrombus
16
imaging laa
16
laa
11
transesophageal echocardiography
8
left atrial
8
atrial appendage
8
thrombus
8
thrombus swine
8
detecting laa
8
laa repeated
8

Similar Publications

Background: Unlike non-rheumatic atrial fibrillation (AF), where left atrial thrombus (LAT) is predominantly confined to the left atrial appendage (LAA), a significant proportion of LAT in rheumatic AF occurs within the left atrial cavity (LAC). However, LAC thrombosis in rheumatic AF has not been extensively studied. This study aimed to evaluate the prevalence of LAT and its subtypes and identify potential predictors of LAT.

View Article and Find Full Text PDF

Background: Left ventricular unloading is needed in patients on extracorporeal life support (ECLS) with severely impaired left ventricular contractility to avoid stasis and pulmonary congestion, and to promote LV recovery. The presence of thrombi in the LV precludes the use of conventional active unloading methods such as transaortic microaxial pumps or apical LV vents. We describe placement of a vent cannula via the left atrial appendage (LAA) as a useful bailout option.

View Article and Find Full Text PDF

Morphing the left atrium geometry: The role of the pulmonary veins on flow patterns and thrombus formation.

Comput Biol Med

January 2025

Departamento de Ingeniería Energética, Universidad Politécnica de Madrid, Avda. de Ramiro de Maeztu 7, Madrid, 28040, Spain. Electronic address:

Background: Despite the significant advances made in the field of computational fluid dynamics (CFD) to simulate the left atrium (LA) in atrial fibrillation (AF) conditions, the connection between atrial structure, flow dynamics, and blood stagnation in the left atrial appendage (LAA) remains unclear. Deepening our understanding of this relationship would have important clinical implications, as the thrombi formed within the LAA are one of the main causes of stroke.

Aim: To highlight and better understand the fundamental role of the PV orientation in forming atrial flow patterns and systematically quantifying its effect on blood stasis within the LAA.

View Article and Find Full Text PDF
Article Synopsis
  • Left atrial appendage closure (LAAC) is an alternative for patients with nonvalvular atrial fibrillation (NVAF) to avoid oral anticoagulants, but issues like incomplete device endothelialization (IDE) can lead to complications.
  • A study compared the endothelialization rates after implantation of two devices: the Watchman plug and the LACBES occluder, in 201 patients, using cardiac computed tomography angiography (CCTA).
  • Results showed that the LACBES occluder had a higher IDE rate at both 3 and 6 months compared to the Watchman device, indicating that it takes longer for the LACBES to achieve complete endothelialization after LAAC.
View Article and Find Full Text PDF
Article Synopsis
  • Stroke is a significant risk following transcatheter aortic valve implantation (TAVI), and this study aimed to determine if patterns from CT scans can predict stroke or transient ischemic attacks (TIA) in patients with severe aortic stenosis.
  • The research included 124 patients and found that stroke/TIA occurred in 9.6% of cases; late-phase filling defects (FDs) in the left atrial appendage (LAA) were significantly linked to stroke, while early-phase FDs showed some correlation, particularly with lower density values.
  • The study concluded that ongoing LAA filling defects observed in late-phase CT scans and certain density criteria from early-phase scans may enhance the assessment of stroke risk, suggesting a
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!