Background: Point of care (POC) devices measuring the international normalized ratio (INR) are accurate for patients with stable disease, but their efficiency has not been prospectively assessed during the "bridging period" when patients are receiving a low molecular weight heparin (LMWH) on top of a vitamin K antagonist (VKA) until the target INR is reached.

Methods: 188 dual INR measurement using the POC (INR(POC)) and the laboratory (INR(lab)) at the same time were consecutively determined : 69 in patients receiving LMWH+VKA (bridging group) and 119 in patients receiving only a VKA (control group). INRpoc was compared to INR(lab).

Results: Test strip failure rate was higher in the bridging group than in the control group (29% vs 4%; p<0,001). In successful tests, POC accuracy was not modified by LMWH administration: the correlation coefficients between POC and lab INR values for the bridging group and the control group were 0,81 and 0,87 respectively, and the relative measure of divergence (RMD=INR(lab) - INR(poc)/INR(lab)) was lower in the bridging group than in the control group (4+/-7% vs 10+/-14%; p=0,02). Finally, clinically relevant agreement between POC and laboratory was of 90% in the bridging group and 92.1% in the control group (p=0.6).

Conclusion: With the POC used (INRatio), in patients receiving LMWH when the POC gives a result, it is as accurate as in patients not receiving a LMWH.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.thromres.2009.09.004DOI Listing

Publication Analysis

Top Keywords

patients receiving
16
international normalized
8
normalized ratio
8
receiving low
8
low molecular
8
molecular weight
8
weight heparin
8
bridging group
8
control group
8
patients
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!