A gynaecological applicator consisting of a metallic intra-uterine tube with a plastic vaginal applicator and an HDR Ir-192 source have been simulated with MCNP5 (Monte Carlo code). A solid phantom has been designed to perform measurements around the applicator with radiochromic films. The isodose curves obtained are compared with curves calculated with the F4MESH tally of MCNP5 with a good agreement. A pinpoint ionization chamber has been used to evaluate dose at some reference points.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.apradiso.2009.10.018DOI Listing