Mapping and isolation of the pulmonary veins using the PVAC catheter.

Pacing Clin Electrophysiol

Department of Cardiology, St-Jan Hospital Bruges, Ruddershove 10, Bruges, Belgium.

Published: February 2010

Objectives: We aimed to investigate the feasibility, efficacy, and safety of the pulmonary vein ablation catheter (PVAC) catheter (a novel multielectrode catheter using duty-cycled bipolar and unipolar radiofrequency energy, Medtronic, Minneapolis, MN, USA) to completely isolate the pulmonary veins (PVs).

Methods: Twenty-seven patients (60 +/- 8 years) with paroxysmal atrial fibrillation (AF) underwent PV isolation with the PVAC catheter. PVAC was used for both mapping and isolation of the PVs (PVAC-guided ablation). After PVAC ablation, presence/absence of PV potentials (PVP) was verified using a conventional circular mapping catheter. In case of residual PVP on the circular catheter, PVAC ablation was continued.

Results: After PVAC-guided ablation 99 of 106 PVs (93%) and 21 of 27 patients (78%) were proven to be isolated. Failure to isolate was due to a mapping failure in four right-sided PVs and a true ablation failure in three right-sided PVs. After continued PVAC ablation, 103 of 106 PVs (97%) and 25 of 27 patients (93%) were shown to be isolated. The total procedural time from femoral vein access to complete catheter withdrawal was 176 +/- 25 minutes. The actual dwelling-time of the PVAC within the left atrium was 102 +/- 37 minutes. Esophageal T degrees rise to >38.5 degrees occurred in nine of 19 monitored patients (47%).

Conclusions: (1) PVAC-guided ablation (i.e., mapping and ablation with a single catheter) results in isolation of all PVs in 73% of the patients. (2) An additional circular mapping catheter is required to increase complete isolation rate to 93% of the patients. (3) Given the esophageal T degrees rise in almost 50% of patients, safety precautions are needed.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1540-8159.2009.02609.xDOI Listing

Publication Analysis

Top Keywords

pvac catheter
12
catheter pvac
12
pvac-guided ablation
12
pvac ablation
12
catheter
10
ablation
9
mapping isolation
8
pulmonary veins
8
pvac
8
isolation pvs
8

Similar Publications

Safety of different therapies for atrial fibrillation: a network meta-analysis.

Curr Probl Cardiol

November 2024

Department of cardiovasccular, The University-Town Hospital of Chongqing Medical University, Chongqing, PR China. Electronic address:

Objective: Most published studies have aimed to compare the effectiveness of different treatment strategies for atrial fibrillation (AF), while few articles have comprehensively compared the safety of therapeutic measures.The aim of the article was to assess the safety of different therapeutic measures (different ablation techniques, antiarrhythmic drugs and surgery) in patients with AF.

Method: A comprehensive and systematic search was undertaken across various databases, namely PubMed, Embase, Cochrane Library, and Web of Science, with the aim of identifying pertinent randomized controlled trials (RCTs) that delve into the safety aspects of diverse atrial fibrillation treatment strategies.

View Article and Find Full Text PDF

Background: There are no randomized studies comparing the maintenance of sinus rhythm after catheter ablation (CA) concerning treatment with antiarrhythmic drugs (AA) in elderly patients with paroxysmal atrial fibrillation (AF).

Objectives: To compare the clinical results of pulmonary vein (PV) isolation with the second-generation PVAC Gold catheter against AA treatment in elderly people with recurrent symptomatic paroxysmal AF, refractory to at least one AA, and without structural heart disease.

Methods: Sixty patients with paroxysmal AF ≥ 65 years old were randomized to two forms of treatment: group 1: CA and group 2: AA drugs.

View Article and Find Full Text PDF

Purpose: The second-generation multi-electrode catheter, PVAC Gold, was designed to improve the safe delivery of phased radiofrequency energy using a "single shot" approach for pulmonary vein isolation (PVI), while retaining efficacy. This large registry presents long-term performance in a daily practice setting.

Methods: A total of 1011 patients undergoing first time ablation for atrial fibrillation (AF) using PVAC Gold were included, 639 patients with PVI for paroxysmal AF (PAF PVI) and 372 patients with persistent or long-standing persistent AF, divided into 175 patients receiving PVI only (PersAF PVI) and 197 patients receiving PVI with additional substrate ablation (PersAF PVI +).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!