Evaluation of monkey intraocular pressure by rebound tonometer.

Mol Vis

Ophthalmic Laboratories & Department of Ophthalmology, West China Hospital, Sichuan University, Chengdu, China.

Published: October 2009

Purpose: To evaluate the usefulness of the TonoVet rebound tonometer in measuring intraocular pressure (IOP) of monkeys.

Methods: The accuracy of the TonoVet rebound tonometer was determined in cannulated eyes of anesthetized rhesus monkeys where IOP was controlled by adjusting the height of a connected perfusate reservoir. To assess the applicability of the equipment through in vivo studies, the diurnal fluctuation of IOP and effects of IOP-lowering compounds were evaluated in monkeys.

Results: IOP readings generated by the TonoVet tonometer correlated very well with the actual pressure in the cannulated monkey eye. The linear correlation had a slope of 0.922+/-0.014 (mean+/-SEM, n=4), a y-intercept of 3.04+/-0.61, and a correlation coefficient of r(2)=0.97. Using this method, diurnal IOP fluctuation of the rhesus monkey was demonstrated. The tonometer was also able to detect IOP changes induced by pharmacologically active compounds. A single topical ocular instillation (15 microg) of the rho kinase inhibitor, H1152, produced a 5-6 mmHg reduction (p<0.001) in IOP, lasting at least 4 h. In addition, topical administration of Travatan, a prostaglandin agonist, induced a small transient IOP increase (1.1 mmHg versus vehicle control; p=0.26) at 2 h after treatment followed by a pressure reduction at 23 h (-2.4 mmHg; p<0.05). Multiple daily dosing with the drug produced a persistent IOP-lowering effect. Three consecutive days of Travatan treatment produced ocular hypotension of -2.0 to -2.2 mmHg (p<0.05) the following day.

Conclusions: The rebound tonometer was easy to use and accurately measured IOP in the rhesus monkey eye.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2773735PMC

Publication Analysis

Top Keywords

rebound tonometer
12
intraocular pressure
8
tonovet rebound
8
iop
6
tonometer
5
evaluation monkey
4
monkey intraocular
4
pressure rebound
4
tonometer purpose
4
purpose evaluate
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!