Introduction: Accurate atrial arrhythmia discrimination is important for dual chamber pacemakers and defibrillators. The aim was to assess the accuracy of atrial arrhythmia recording using modern devices and relate this to atrial tip-to-ring (TTR) distance.
Methods: One hundred eighty-two patients (72 + or - 9 years, 55% male) with paroxysmal atrial fibrillation were enrolled and were included in the study if they had an atrial fibrillation (AF) burden of 1-50% during a monitoring phase. Seventy-nine patients fulfilled these criteria and were followed for at least 5 months. Electrodes were classified as having short (<10 mm), medium (10-12), or long (13-18) atrial TTR spacing.
Results: Two thousand eight hundred eighty-three detailed onset reports were analyzed; 730 (25%) demonstrated aberrant sensing. Six percent were due to farfield R wave oversensing (FFRWO) and 19% due to undersensing, sometimes occurring in the same patient and study phase. FFRWO was significantly reduced with short TTR electrodes (P < 0.05). Undersensing due to sensitivity fallout was 18% (short), 24% (medium), and 17% (long) (P = ns). Undersensing due to pacemaker blanking was 11% (short), 11% (medium), and 12% (long) (P = ns). Active fixation electrodes did not show any difference from passive fixation.
Conclusion: Atrial electrodes with a short TTR (<10 mm) significantly reduce FFRWO without increasing undersensing and should be used routinely in patients with paroxysmal atrial tachyarrhythmias. However, 20% of atrial tachyarrythmia episodes were incorrectly classified as terminated by these modern devices due to undersensing. Clinicians should be wary of using device-derived endpoints that rely on AF episode number or duration as these may be falsely increased or reduced, respectively.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/j.1540-8159.2009.02576.x | DOI Listing |
J Cardiovasc Electrophysiol
January 2025
Douala Gyneco-obstetric and Pediatric Hospital/University of Douala, Douala, Cameroon.
Cureus
December 2024
Cardiovascular Surgery, Kawasaki Municipal Hospital, Kawasaki, JPN.
A 40-year-old male visited our clinic for cardiac evaluation. He had palpitations for several years, but the reason was unknown. Transthoracic echocardiography revealed a hyperechoic ribbon-shaped structure that moved vigorously in the right atrium.
View Article and Find Full Text PDFCardiovasc Eng Technol
January 2025
Department of Biomedical Engineering, Northwestern University, Chicago, IL, USA.
Purpose: This study explores the use of heart rate variability (HRV) analysis, a noninvasive technique for assessing the autonomic nervous system, by applying nonlinear dynamics and chaos theory to detect chaotic behavior in RR intervals and assess cardiovascular health.
Methods: Employing the "System Analysis of Heart Rate Dynamics" (SADR) program, this research combines chaos analysis with the short-time Fourier transform to assess nonlinear dynamic parameters in HRV. It includes constructing phase portraits in Takens space and calculating measures of chaos to identify deterministic chaos indicators.
Sci Rep
January 2025
Department of Nephrology, The Third Affiliated Hospital of Sun Yat-sen University, Guangzhou, China.
Systemic inflammation plays a crucial role in the pathogenesis and prognosis of diabetes and cardiovascular diseases. System inflammation response index (SIRI), is an emerging biomarker designed to assess the extent of systemic inflammation. We aimed to delineate the prognostic significance of SIRI in patients with both AF and type 2 diabetes mellitus (T2DM).
View Article and Find Full Text PDFBMJ Open
January 2025
Department of Medical Research Bærum Hospital, Vestre Viken Hospital Trust, Gjettum, Norway
Objectives: The objective of this study was to assess the quality of ECG recordings and the concordance between the automatic detection of cardiac arrhythmia episodes by a patch ECG and an insertable cardiac monitor.
Design: Prospective cohort study.
Setting And Participants: Endurance athletes diagnosed with paroxysmal atrial fibrillation (AF) and no other relevant comorbidities participating in a randomised controlled trial on the effects of training adaption.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!