Background: Implementing tight glycemic control (TGC) in intensive care unit (ICU) patients requires accurate blood glucose (BG) monitoring. We evaluated the performance of two commercially available bedside glucometers, Accu-Chek and HemoCue, in patients admitted to the ICU and in whom TGC was applied.

Methods: Thirty-seven adult ICU patients were prospectively included. During 48 hours, BG was determined simultaneously on the same arterial blood sample using the two point-of-care testing (POCT) glucometers as compared with the standard technique. Data of 452 paired measurements were analyzed using linear regression, Clark error grid analysis (EGA), the method of Bland-Altman, and the GLYCENSIT procedure.

Results: Both tested glucometers showed satisfactory results when evaluated with linear regression and EGA. Correlation coefficients were above 0.9, and 100% of all the glucose readings were within the safe zones A and B using EGA. However, when applying more appropriate tests, both sensors failed to provide sufficient accuracy in the setting of TGC in ICU patients. The Hemocue revealed a bias of >10 mg/dl with a trend to systematically overestimate the actual BG value. The bias for the Accu-Chek was 6 mg/dl with wide limits of agreement and a variable over- and underestimation of the actual BG value depending on the level of BG (hypo-, normo-, or hyperglycemia).

Conclusions: When TGC is implemented in ICU practice, caution is warranted when adjusting insulin rates based only on BG readings obtained by the tested glucometers. ICU practitioners should weigh the advantages and disadvantages of such devices: a greater bias but with a more predictable error and measurement behavior versus a somewhat lower bias but with an unpredictable direction of the difference.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2769815PMC
http://dx.doi.org/10.1177/193229680800200603DOI Listing

Publication Analysis

Top Keywords

icu patients
12
blood glucose
8
arterial blood
8
intensive care
8
care unit
8
tight glycemic
8
glycemic control
8
linear regression
8
tested glucometers
8
icu
6

Similar Publications

Importance: This study addresses the characteristics, kidney replacement therapy (KRT) modalities, and outcomes in children diagnosed with crush syndrome following an earthquake in Turkey.

Objective: To analyze the associations of different KRT modalities with long-term dialysis dependency and length of stay (LOS) in the pediatric intensive care unit (PICU).

Design, Setting, And Participants: This multicenter, prospective, and retrospective cohort study was conducted across 20 PICUs in Turkey.

View Article and Find Full Text PDF

Background: Machine perfusion (MP), including hypothermic oxygenated machine perfusion (HOPE), dual HOPE, normothermic machine perfusion (NMP), NMP ischemia-free liver transplantation (NMP-ILT), and controlled oxygenated rewarming (COR), is increasingly being investigated to improve liver graft quality from extended criteria donors and donors after circulatory death and expand the donor pool. This network meta-analysis investigates the comparative efficacy and safety of various liver MP strategies versus traditional static cold storage (SCS).

Methods: We searched PubMed, Scopus, Web of Science, and Cochrane Controlled Register of Trials for randomized controlled trials (RCTs) comparing liver transplantation (LT) outcomes between SCS and MP techniques.

View Article and Find Full Text PDF

Background: First Nations patients often experience poorer health outcomes than non-First Nations patients. Despite emergency triage primarily focusing on severity, implying comparable outcomes for patients in the same triage group regardless of demographics, the precision of triage for First-Nations Australians may be undermined by multiple factors, although research in this area is scarce.

Objective: To compare admission rates, service utilisation and mortality for First Nations and non-First Nations patients, based on their triage categories.

View Article and Find Full Text PDF

Herein, we report the cases of two patients with hemolysis, elevated liver enzymes, and low platelets syndrome who underwent emergent Cesarean sections that were complicated by massive hemorrhage due to undiagnosed hepatic rupture. Intraoperative General Surgery team intervention, early activation of massive transfusion protocol, hemostatic resuscitation, and transfer to ICU resulted in the survival of both patients.

View Article and Find Full Text PDF

Background: Rhabdomyolysis (RM) frequently gives rise to diverse complications, ultimately leading to an unfavorable prognosis for patients. Consequently, there is a pressing need for early prediction of survival rates among RM patients, yet reliable and effective predictive models are currently scarce.

Methods: All data utilized in this study were sourced from the MIMIC-IV database.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!