Volumetric high-intensity focused ultrasound (HIFU) guided by multiplane magnetic resonance (MR) thermometry has been shown to be a safe and efficient method to thermally ablate large tissue volumes. However, the induced temperature rise and thermal lesions show significant variability, depending on exposure parameters, such as power and timing, as well as unknown tissue parameters. In this study, a simple and robust feedback-control method that relies on rapid MR thermometry to control the HIFU exposure during heating is introduced. The binary feedback algorithm adjusts the durations of the concentric ablation circles within the target volume to reach an optimal temperature. The efficacy of the binary feedback control was evaluated by performing 90 ablations in vivo and comparing the results with simulations. Feedback control of the sonications improved the reproducibility of the induced lesion size. The standard deviation of the diameter was reduced by factors of 1.9, 7.2, 5.0, and 3.4 for 4-, 8-, 12-, and 16-mm lesions, respectively. Energy efficiency was also improved, as the binary feedback method required less energy to create the desired lesion. These results show that binary feedback improves the quality of volumetric ablation by consistently producing thermal lesions of expected size while reducing the required energy as well.

Download full-text PDF

Source
http://dx.doi.org/10.1109/TBME.2009.2034636DOI Listing

Publication Analysis

Top Keywords

binary feedback
20
feedback control
12
thermal lesions
8
required energy
8
feedback
6
binary
5
improved volumetric
4
volumetric mr-hifu
4
mr-hifu ablation
4
ablation robust
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!