Purpose: Previous studies have shown the ability of rubidium-82 ((82)Rb) positron emission tomography (PET) imaging to quantitatively measure myocardial blood flow (MBF), many of which are performed using two-dimensional (2D) imaging. Three-dimensional (3D) imaging provides increased sensitivity and may result in decreased costs owing to a reduction in the required injected activity of radiotracer. This study compares 2D and 3D (82)Rb PET MBF results obtained in the same imaging session.

Methods: Three-dimensional and 2D (82)Rb perfusion imaging was performed in canines on a GE Discovery LS PET/CT scanner at rest and during hyperemia in stunned and infarcted tissue. MBF (ml/min/g) was determined using a 1-compartment model and an extraction correction of the uptake rate and analyzed using a standard 17-segment model.

Results: A strong, significant correlation was present (rho = 0.95, P<0.0001). Average 3D MBF values were slightly lower at rest and higher during stress versus 2D. MBF results in normal, stunned, and infarcted tissue differed by 7% on average and significant increases in MBF from rest to hyperemia were noted with both the techniques.

Conclusion: These results imply that MBF results obtained in 3D are comparable with traditional 2D imaging. Therefore, it may be possible to use 3D imaging with lower administered activity, helping to reduce costs and patient dose without compromising quantitative information.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MNM.0b013e328332b359DOI Listing

Publication Analysis

Top Keywords

myocardial blood
8
blood flow
8
stunned infarcted
8
imaging
6
versus dynamic
4
82rb
4
dynamic 82rb
4
82rb myocardial
4
flow imaging
4
imaging canine
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!