Design, Fabrication and Testing of an Insertable Double-Imaging-Region Gradient Coil.

Concepts Magn Reson Part B Magn Reson Eng

Utah Center for Advanced Imaging Research, Department of Radiology, University of Utah, Salt Lake City, Utah, USA.

Published: April 2009

We have constructed a small-bore insertable gradient coil with two linear gradient imaging regions and interfaced it with an MRI scanner. We have also constructed an RF system capable of transmitting or receiving in both regions simultaneously.Designs for conductor placement for two-region X-, Y- and Z-gradient coils were optimized by simulated annealing. Wire patterns for each axis were chosen that gave low inductance, reasonable homogeneity over a large imaging volume and high efficiency (gradient field per-unit-current).Imaging was performed on a Siemens 3T TIM Trio scanner equipped with three additional gradient amplifier channels and a second RF/gradient array controller. Phantoms were placed in the two imaging regions as well as the central non-imaging region to test gradient homogeneity and crosstalk between regions. Images acquired simultaneously in the two regions showed very little signal crosstalk between imaging regions and even less signal from the central, non-imaging region.When combined with an overlapping single-region gradient insert, extended field-of-view (FOV) imaging will be possible without moving the table or the subject and without increasing nerve stimulation. Construction and testing of a two-region gradient coil insert is a necessary intermediate step as a proof of concept for an extended field of view, contiguous, three-region human-sized gradient system.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2760823PMC
http://dx.doi.org/10.1002/cmr.b.20138DOI Listing

Publication Analysis

Top Keywords

gradient coil
12
imaging regions
12
gradient
9
central non-imaging
8
regions signal
8
regions
6
imaging
5
design fabrication
4
fabrication testing
4
testing insertable
4

Similar Publications

Fast Hadamard-Encoded 7T Spectroscopic Imaging of Human Brain.

Tomography

January 2025

NextGen Precision Health, Department of Radiology, University of Missouri Columbia, 1030 Hitt Street, Columbia, MO 65201, USA.

: The increased SNR available at 7T combined with fast readout trajectories enables accelerated spectroscopic imaging acquisitions for clinical applications. In this report, we evaluate the performance of a Hadamard slice encoding strategy with a 2D rosette trajectory for multi-slice fast spectroscopic imaging at 7T. : Moderate-TE (~40 ms) spin echo and J-refocused polarization transfer sequences were acquired with simultaneous Hadamard multi-slice excitations and rosette in-plane encoding.

View Article and Find Full Text PDF

Background: Quantitative Susceptibility Mapping (QSM) provides a non-invasive post-processing method to investigate alterations in magnetic susceptibility (χ), reflecting iron content within brain regions implicated in neurodegenerative diseases (NDDs).

Purpose: To investigate alterations in thalamic χ in patients with NDDs using QSM.

Study Type: Systematic review and meta-analysis.

View Article and Find Full Text PDF

Magnetic nanoparticle hyperthermia (MNH) emerges as a promising therapeutic strategy for cancer treatment, leveraging alternating magnetic fields (AMFs) to induce localized heating through magnetic nanoparticles (MNPs). However, the interaction of AMFs with biological tissues leads to non-specific heating caused by eddy currents, triggering thermoregulatory responses and complex thermal gradients throughout the body of the patient. While previous studies have implemented the Atkinson-Brezovich limit to mitigate potential harm, recent research underscores discrepancies between this threshold and clinical outcomes, necessitating a re-evaluation of this safety limit.

View Article and Find Full Text PDF

Exploring in vivo human brain metabolism at 10.5 T: Initial insights from MR spectroscopic imaging.

Neuroimage

January 2025

Center for Magnetic Resonance Research, Department of Radiology, University of Minnesota, Minneapolis, USA. Electronic address:

Introduction: Ultra-high-field magnetic resonance (MR) systems (7 T and 9.4 T) offer the ability to probe human brain metabolism with enhanced precision. Here, we present the preliminary findings from 3D MR spectroscopic imaging (MRSI) of the human brain conducted with the world's first 10.

View Article and Find Full Text PDF

Background: There is evidence that iron metabolism may play a role in the underlying pathophysiological mechanism of migraine. Studies using (=1/ ) relaxometry, a common MRI-based iron mapping technique, have reported increased values in various brain structures of migraineurs, indicating iron accumulation compared to healthy controls.

Purpose: To investigate whether there are short-term changes in during a migraine attack.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!