Background: To determine the preferable method for self-injecting insulin, we compared the handling, safety, and dose accuracy of a conventional disposable syringe and vial with FlexPen (Novo Nordisk A/S, Bagsvaerd, Denmark), a prefilled pen.
Methods: Insulin therapy-naive healthcare professionals (HCPs) (n = 30), unfamiliar with insulin delivery, injected 10 U of insulin into a sponge pad using either a syringe and vial or the FlexPen, both with 30-gauge 8-mm needles, on day 1. The following day, they used the alternative method. They evaluated the handling of the two methods on device-specific questionnaires and compared overall preference on a third questionnaire. To evaluate dose accuracy, 30 insulin therapy-experienced HCPs and 20 insulin therapy-naive HCPs were asked to deliver 10 U of insulin using each method, and the amount discharged was weighed.
Results: FlexPen was rated easier to use and overall more preferable than the syringe and vial by insulin therapy-naive HCPs (P < 0.001). The pen device was more accurate than the syringe and vial when used by experienced HCPs (mean +/- SD dose delivered, 9.91 +/- 0.11 U vs. 9.82 +/- 0.25 U, respectively; P < 0.001) and by insulin therapy-naive HCPs (9.91 +/- 0.12 U vs. 9.74 +/- 0.85 U; P < 0.001).
Conclusions: Insulin therapy-naive HCPs found FlexPen easier to handle and preferable to use compared to a conventional syringe and vial. Both insulin therapy-experienced and -naive HCPs were able to deliver insulin significantly more accurately with the FlexPen than with a syringe and vial (P < 0.001).
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1089/dia.2009.0006 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!