A detailed investigation into the performance of two flow injection-chemiluminescence (FI-CL) manifolds (with and without a preconcentration column) for the determination of sub-nanomolar dissolved iron (Fe(II)+Fe(III)), following the reduction of Fe(III) by sulphite, in seawater is described. Kinetic experiments were conducted to examine the efficiency of reduction of inorganic Fe(III) with sulphite under different conditions and a rigorous study of the potential interference caused by other transition metals present in seawater was conducted. Using 100microM concentrations of sulphite a reduction time of 4h was sufficient to quantitatively reduce Fe(III) in seawater. Under optimal conditions, cobalt(II) and vanadium(IV)/(III) were the major positive interferences and strategies for their removal are reported. Specifically, cobalt(II) was masked by the addition of dimethylglyoxime to the luminol solution and vanadium(IV) was removed by passing the sample through an 8-hydroxyquinoline column in a low pH carrier stream. Manganese(II) also interfered by suppression of the CL response but this was not significant at typical open ocean concentrations.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.aca.2009.06.011DOI Listing

Publication Analysis

Top Keywords

dissolved iron
8
feiii sulphite
8
investigation ironiii
4
reduction
4
ironiii reduction
4
reduction trace
4
trace metal
4
metal interferences
4
interferences determination
4
determination dissolved
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?