Background And Objectives: Blood donor enrolment process is frequently based on the sole capillary haemoglobin (Hb) evaluation while platelet donors by apheresis also requires platelet (Plt) count. The 'sole Hb' approach prevents a complete donor evaluation and does not allow Plt donor enrolment. To extend blood counts before donations, we evaluated the performances of a multiparametric counter using capillary blood.

Materials And Methods: The ABX Micros 60 (Micros 60) blood analyzer was employed on capillary blood and compared with venous counts by a reference counter (Coulter AcT 5diff) in a first series of 416 donors and in a second series of 136, after a 3-month period of routine use of this study counter. An average of 50 microl of capillary blood was collected whose 10 microl had been aspirated by Micros 60.

Results: High correlations were found between capillary counts using Micros 60 and venous counts using the reference counter. Mean Plt counts differed of 37 x 10(9)/l less for capillary approach in the first series of comparisons, but decreased to 10 x 10(9)/l less in the second series due to a greater expertise of operators in capillary sampling. All other parameters were accurate and never reached clinical relevance albeit they showed statistically significant differences.

Conclusion: Data on Micros 60 demonstrated that capillary predonation counts may represent a feasible and effective approach to realize an accurate enrolment process of blood and Plt donors.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1423-0410.2009.01256.xDOI Listing

Publication Analysis

Top Keywords

capillary blood
12
donor enrolment
12
capillary
9
blood
8
blood donor
8
enrolment process
8
venous counts
8
counts reference
8
reference counter
8
second series
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!