A life-threatening arrhythmia induced by inappropriate activation of an implantable cardioverter defibrillator.

Europace

Department of Cardiology, Rabin Medical Center, Beilinson Hospital-Petah Tikva, affiliated to the Sackler School of Medicine, Tel Aviv University, 39 Jabotinski Street, Petah Tikva, Israel.

Published: December 2009

We describe an unusual case of near fatal inappropriate implantable cardioverter defibrillator therapies due to atrial oversensing by a newly implanted ventricular lead. Chest X-ray revealed dislodgement of the active fixation lead to the tricuspid annulus area explaining the atrial oversensing and intermittent ventricular therapies.