Objective: The project aimed to design an instrument that can provide visual, quantitative feedback to medical students and help them develop intuitive, tactile-based injection skills.

Methods: The instrument uses force and displacement sensors to find flow impedance, the ratio of the pressure applied to fluid in the syringe to the fluid flow rate, of different tissues. A software program utilizes these flow impedances to identify needle tip location-either joint or tendon. Injection experiments with pork feet were completed to collect flow impedance data when the needle tip was located in joint or tendon. Injections used deionized water and aqueous pullulan solutions to study the instrument's effectiveness when medical therapies of different viscosities are injected. For each solution, at least 13 injections were conducted. The mean and standard deviation of the flow impedances from injections with each solution were obtained.

Results: In all experimental cases, the flow impedance demonstrated by the tendon is a number of times higher than that of the joint cavity: the impedance ratios (tendon to joint cavity) for water, 3 w/w% pullulan, and 15 w/w% pullulan are 74, 62, and 6, respectively.

Conclusions: The results confirmed the concept of using flow impedance to index the needle tip location-whether the tip is positioned in joint or tendon. This instrument's user interface can serve as a training aid for medical students and infrequent operators.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10877-009-9202-zDOI Listing

Publication Analysis

Top Keywords

flow impedance
20
joint tendon
12
flow
8
medical students
8
flow impedances
8
joint cavity
8
w/w% pullulan
8
impedance
6
joint
5
tendon
5

Similar Publications

Article Synopsis
  • The study examines how the rheology (flow behavior) of ingested fluids affects swallowing and the physiological responses during deglutition, specifically comparing xanthan gum (XG) and sodium carboxymethylcellulose gum (CMC) in healthy adults.
  • Results showed that CMC had significantly higher viscosity than XG at higher shear rates (300 s), leading to increased flow resistance during swallowing, indicated by higher intrabolus pressure and altered relaxation times of the upper esophageal sphincter (UES).
  • The findings suggest that the differences in shear viscosity of these fluids affect pharyngeal function during swallowing, highlighting the importance of fluid properties over standardized viscosity levels (IDDSI).
View Article and Find Full Text PDF

A curvature of a finger that bends inwards relative to the other fingers is a common observation during prenatal screening. When the angulation exceeds 10 degrees, it is known as "clinodactyly" and could suggest a variety of underlying issues. Even though it usually remains unnoticed during pregnancy, it may be a sign of serious fetal disease.

View Article and Find Full Text PDF

Solving the surface (electro-)chemical instability and the fading behavior of high voltage cathode materials cycled above 4.3 V vs Li/Li remains a major challenge for the next generation of high energy density Li-ion batteries. Here, we present a facile, environmentally friendly, cost effective and scalable method to address this problem by uniformly fluorinating the surface of cathode materials with a mild fluorinating agent (CHF) using a gas flow-type reactor.

View Article and Find Full Text PDF

Microfluidic impedance flow cytometer leveraging virtual constriction microchannel and its application in leukocyte differential.

Microsyst Nanoeng

December 2024

State Key Laboratory of Transducer Technology, Aerospace Information Research Institute, Chinese Academy of Sciences, Beijing, 100190, People's Republic of China.

Microfluidic impedance flow cytometry has been widely used in leukocyte differential and counting, but it faces a bottleneck due to the trade-off between impedance detection throughput and sensitivity. In this study, a microfluidic impedance flow cytometer based on a virtual constriction microchannel was reported, in which the virtual constriction microchannel was constructed by crossflow of conductive sample and insulated sheath fluids with underneath micro-electrodes for impedance measurements. Compared to conventional mechanical constriction microchannels, this virtual counterpart could effectively avoid direct physical contact between cells and the microchannel walls to maintain high throughputs, and significantly reduce the volume of the impedance detection region for sensitivity improvements.

View Article and Find Full Text PDF

Age and Hypertension Synergize With Dehydration to Cause Renal Frailty in Rats and Predispose Them to Intrinsic Acute Kidney Injury.

Lab Invest

December 2024

Translational Research on Renal and Cardiovascular Diseases (TRECARD), Department of Physiology and Pharmacology, Universidad de Salamanca, Salamanca, Spain; Instituto de Investigación Biomédica de Salamanca (IBSAL), Salamanca, Spain. Electronic address:

Acute kidney frailty (AKF) is a condition of increased susceptibility to acute kidney injury (AKI), an abrupt impairment of renal excretory function potentially leading to severe complications. Prevention of AKI relies on the recognition of risk factors contributing to AKF. At the population level, dehydration constitutes a predisposing factor for AKI.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!