Purpose: The purpose of this study was to evaluate the accuracy of a new version of the ExacTrac X-ray (ETX) system with statistical analysis retrospectively in order to determine the tolerance of systematic components of spatial uncertainties with the Novalis system.
Methods And Materials: Three factors of geometrical accuracy related to the ETX system were evaluated by phantom studies. First, location dependency of the detection ability of the infrared system was evaluated. Second, accuracy of the automated calculation by the image fusion algorithm in the patient registration software was evaluated. Third, deviation of the coordinate scale between the ETX isocenter and the mechanical isocenter was evaluated. From the values of these examinations and clinical experiences, the total spatial uncertainty with the Novalis system was evaluated.
Results: As to the location dependency of the detection ability of the infrared system, the detection errors between the actual position and the detected position were 1% in translation shift and 0.1 degrees in rotational angle, respectively. As to the accuracy of patient verification software, the repeatability and the coincidence of the calculation value by image fusion were good when the contrast of the X-ray image was high. The deviation of coordinates between the ETX isocenter and the mechanical isocenter was 0.313 +/- 0.024 mm, in a suitable procedure.
Conclusions: The spatial uncertainty will be less than 2 mm when suitable treatment planning, optimal patient setup, and daily quality assurance for the Novalis system are achieved in the routine workload.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ijrobp.2009.02.080 | DOI Listing |
Phys Med
December 2024
Department of Radiation Oncology, Vanderbilt University Medical Center, Nashville, TN 37232, USA. Electronic address:
Background: Stereotactic radiosurgery (SRS) is an effective treatment for essential tremor (ET) and Parkinsonian tremor (PT). However, current treatment methods can be time-consuming and expose patients to unnecessary radiation. This study investigates the use of the Varian TrueBeam system with FFF beam modes as a potential solution for these issues.
View Article and Find Full Text PDFAsian Pac J Cancer Prev
August 2024
Department of Physics, School of Advanced Sciences, Vellore Institute of Technology, Vellore, Tamil Nadu, India.
Objective: This study addresses challenges in delivering high radiation doses and managing organ motion in Stereotactic Body Radiation Therapy (SBRT) for thoracic and abdominal cancer. It evaluates Varian's Real Time Position Management (RPM) system's infrared camera sensitivity during crucial Four-Dimensional computed tomography (4D-CT) scans for planning and treatment. The analysis includes CT simulator, LINAC (Novalis Tx and TrueBeam STx).
View Article and Find Full Text PDFCancer Radiother
April 2024
Département de radiothérapie, centre Jean-Perrin, 58, rue Montalembert, 63011 Clermont-Ferrand cedex 1, France; Institut Pascal, CNRS, Clermont-Auvergne INP, université Clermont-Auvergne, Clermont-Ferrand, France; Inserm U1240 IMoST, université Clermont-Auvergne, Clermont-Ferrand, France. Electronic address:
Purpose: Radiosurgery for the treatment of trigeminal neuralgia delivers a very high dose in a single fraction, over a few millimeters, at a single isocenter placed along the nerve. We present here the different steps that have been performed to validate small beams by conical collimators, and report the clinical results of the first patients treated on Novalis Tx®, frameless.
Material And Methods: First, the geometric accuracy of 4 and 6mm conical collimators was evaluated using Winston-Lutz tests; then dosimetric data acquisition was performed using high spatial resolution detectors (PTW 60019 microdiamond and a PTW 60017 E-diode).
Surg Neurol Int
December 2023
Department of Neurosurgery, Yokohama Rosai Hospital, Yokohama, Japan.
Background: Skull base meningiomas are often difficult to remove completely with preserved nerve function and may require radiation therapy. However, the Gamma Knife is unsuitable for large tumor volume or the optic nerve, which is difficult to identify on imaging. We report the results of stereotactic radiotherapy with HybridArc using Novalis STx for skull base meningiomas.
View Article and Find Full Text PDFAnticancer Res
April 2023
Department of Radiation Oncology, Osaka University Graduate School of Medicine, Suita, Japan.
Background/aim: Stereotactic body radiotherapy for prostate cancer using CyberKnife with circular cone requires a long treatment time. Raster scanning intensity modulated radiotherapy (RS-IMRT) has a potential of improving treatment efficacy, introducing shorter treatment time, better target dose uniformity, and lower organ at risk (OAR) dose. The purpose of the study was to develop a fluence optimization system for RS-IMRT.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!