Purpose: To evaluate the performance of a self-contained, battery-driven, vacuum-assisted breast biopsy (VABB) system for the sampling of clustered breast microcalcifications and masses under stereotactic guidance.

Methods And Materials: A total of 144 patients (median age: 56 years; range: 21-87 years) in four European breast centers underwent percutaneous 9-gauge (G), stereotactic-guided VABB. The median lesion size was 11 mm (range 2-60 mm). Patients were biopsied in the prone (n=125) or upright position (n=19). All patients were followed up for at least 24 months.

Results: The stereotactic procedure was successful in 142 (98.6%) of 144 cases, with two cases cancelled due to either severe patient motion (one case) or failure to detect faint calcifications (one case). A median of 12 specimens per procedure was obtained. In 39 cases (27.5%), the suspicious lesion could no longer be detected mammographically after the biopsy procedure. The histological diagnosis was malignancy in 45 (31.7%) cases. One case of atypical ductal hyperplasia diagnosed preoperatively was upgraded to ductal carcinoma in situ (DCIS) at operation, giving an overall sensitivity of 97.7% for the vacuum-assisted biopsy procedure. In two cases where DCIS was diagnosed at vacuum-assisted biopsy, the malignant tissue was apparently completely removed and could no longer be found at operation. No serious complications occurred. During the follow-up period, no breast cancers appeared at the location of biopsy. Six patients dropped out during the follow-up period.

Conclusion: The self-contained, vacuum-assisted biopsy device is well suited for stereotactically guided breast biopsies, having demonstrated excellent sensitivity and specificity in the preoperative workup of mammographically detected breast lesions after 2 years of follow-up.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clinimag.2008.12.010DOI Listing

Publication Analysis

Top Keywords

vacuum-assisted biopsy
12
stereotactically guided
8
breast biopsy
8
self-contained vacuum-assisted
8
procedure cases
8
biopsy procedure
8
breast
7
biopsy
7
vacuum-assisted
5
cases
5

Similar Publications

Assessing Malignant Risk in B3 Breast Lesions: Clinical Insights and Implications.

J Clin Med

December 2024

Multidisciplinary Breast Centre, Dipartimento Scienze della Salute della Donna e del Bambino e di Sanità Pubblica, Fondazione Policlinico Universitario Agostino Gemelli IRCCS, 00168 Rome, Italy.

: B3 breast lesions, characterized by uncertain malignant potential, pose a significant challenge for clinicians. With the increasing use of preoperative biopsies, there is a need for careful management strategies, including watchful waiting, vacuum-assisted excision (VAE), and surgery. This study aims to assess the concordance between preoperative biopsy findings and postoperative histology, with a focus on evaluating the positive predictive value (PPV) for malignancy in B3 lesions.

View Article and Find Full Text PDF

Granulomatous mastitis (GM) poses challenges in diagnosis and treatment due to its similarities with other breast diseases like cancer. The comparative study evaluated the efficacy and safety of a vacuum-assisted biopsy device with minimally invasive excisions compared to traditional wide local excisions. The vacuum-assisted biopsy device technique offers benefits such as precise tissue removal, reduced damage to healthy tissue, shorter surgery and recovery times, and lower postoperative complication risks.

View Article and Find Full Text PDF

Effect of needle size on outcomes of vacuum-assisted excision of breast lesions. A randomized controlled trial.

Eur J Radiol

December 2024

Department of Radiology, Karolinska University Hospital, Stockholm, Sweden; Department of Oncology and Pathology, Karolinska Institutet, Stockholm, Sweden.

Background: Utilizing a larger needle-size instead of a smaller one in vacuum-assisted excision of breast lesions might enhance the effectiveness of the method. We conducted a clinical trial to investigate the effects of needle size 7G compared to 10G regarding excision completeness and procedural efficiency.

Materials And Methods: In this prospective, single-blinded, randomized clinical trial, the patients were enrolled between November 2019 and August 2022.

View Article and Find Full Text PDF

Diagnostic performance of MRI-guided vacuum-assisted breast biopsy (VABB): an essential but still underused technique.

Breast Cancer Res Treat

December 2024

Breast Radiology Department, Fondazione IRCCS Istituto Nazionale Dei Tumori Di Milano, Via Giacomo Venezian 1, 20133, Milan, Italy.

Background: Magnetic resonance imaging (MRI)-guided vacuum-assisted breast biopsy (VABB) is an increasingly requested procedure, but it implies training and experience both in its execution and in determining radiological-pathological concordance and is therefore performed in dedicated breast centers. The purpose of this study is to evaluate the diagnostic performance of MRI-guided vacuum-assisted biopsy and to determine the upgrade rate after surgery or follow-up.

Methods: We retrospectively evaluated all consecutive patients with suspicious MRI findings without corresponding mammographic and ultrasonographic findings who underwent MRI-guided vacuum-assisted breast biopsy (VABB) at our Institution from November 2020 to March 2023.

View Article and Find Full Text PDF

Fibromatosis-like metaplastic carcinoma (FLMC) is a rare subtype of metaplastic carcinoma of the breast. Diagnosing this entity poses significant challenges, particularly in core biopsies due to limited sampling and overlap with benign spindle cell lesions such as nodular fasciitis and fibromatosis. We present an example of FLMC in an asymptomatic middle-aged woman.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!