[Association between stent fracture and restenosis after drug-eluting stent implantation].

Zhonghua Xin Xue Guan Bing Za Zhi

Department of Cardiology, Beijing Anzhen Hospital, Capital Medical University, Beijing 100029, China.

Published: January 2009

Objective: This study was performed to evaluate the relationship between the stent fracture and restenosis after drug-eluting stent implantation.

Methods: The study enrolled 536 patients with angiographies during stenting procedure and follow-up, the patients were divided into DES group (n=397) and BMS group (n=139). The coronary angiography images were analyzed to detect restenosis and stent fracture.

Results: Restenosis rate was significantly lower in DES group (31/397, 7.8%) compared that in BMS group (30/139, 21.6%, P<0.05). Stent fracture (n=5) was found only in DES group and not in BMS group (P<0.05). Restenosis were found in all stent fracture segments. The stent fracture developed at the angulated tortuosity lesions.

Conclusion: Stent fracture is one of the causes of restenosis after drug-eluting stents implantation and related to implantation of long DES stent at the location of angulated tortuosity lesions.