Introduction: This study was designed to test the hypothesis of equivalence in cardiac output (CO) and stroke volume (SV) monitoring capabilities of two devices: non invasive transthoracic bioreactance (NICOM), and a pulse contour analysis (PICCO PC) coupled to transpulmonary thermodilution (PICCO TD).
Methods: We included consecutive patients of a single ICU following cardiac surgery. Continuous minute-by-minute hemodynamic variables obtained from NICOM and PICCO PC were recorded and compared in 20 patients at baseline, during a lung recruitment maneuver (20 cmH2O of PEEP) and following withdrawal of PEEP. PICCO TD measurements were also determined. We evaluated the accuracy of these two technologies at baseline using PICCO TD as reference and we estimated the precision by the fluctuation around the mean value (2SD/mean). Then, we assessed time response, amplitude response and reliability for detecting expected decreases when PEEP was applied. Type I and type II errors were analyzed.
Results: CO values (PICCO TD) ranged from 1.6 to 8.0 L.min-1. At baseline, CO values were comparable for NICOM, PICCO PC and PICCO TD: 5.0 +/- 1.2, 4.7 +/- 1.4 and 4.6 +/- 1.3 L.min.-1, respectively (NS). Limits of agreements with PICCO TD were 1.52 L.min.-1 for NICOM and 1.77 L.min.-1 for PICCO PC, NS. The 95% statistical power gives an equivalence with a threshold of 0.52 L.min.-1 for NICOM vs. PICCO PC. The CO precision was 6 +/- 3% and 6 +/- 5% for NICOM and PICCO PC, respectively, NS. When PEEP was applied, CO was reduced by 33 +/- 12%, 31 +/- 14% and 32 +/- 13%, for NICOM, PICCO PC and PICCO TD, respectively (NS). Time response was 3.2 +/- 0.7 minute for NICOM vs. 2 +/- 0.5 minute for PICCO PC (NS). SV results were comparable to those for CO.
Conclusions: Although limited to 20 patients, this study has enough power to show comparable CO and SV monitoring capabilities of Bioreactance and pulse contour analysis calibrated by transpulmonary thermodilution.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2750178 | PMC |
http://dx.doi.org/10.1186/cc7981 | DOI Listing |
Ann Palliat Med
September 2020
Department of Critical Care Medicine, The Fourth Hospital of Hebei Medical University, Shijiazhuang, China.
Background: The consistency of cardiac output (CO) measured by noninvasive cardiac output monitoring (NICOM), pulse index continuous cardiac output (PiCCO), and ultrasound in the hemodynamic monitoring of critically ill patients was studied. Using the NICOM built-in passive leg raising (PLR) test, stroke volume index variation (∆SVI) was calculated and was used to predict volume responsiveness in patients with circulatory shock (excluding cardiogenic shock).
Methods: Critically ill patients requiring hemodynamic monitoring were admitted during the study period.
J Clin Monit Comput
February 2018
Department of Critical Care Medicine, University of Pittsburgh Medical Center, Pittsburgh, PA, USA.
We compared the similarity of cardiac-output (CO) estimates between available bolus thermodilution pulmonary-artery catheters (PAC), arterial pulse-contour analysis (LiDCOplus, FloTrac and PiCCOplus), and bioreactance (NICOM). Repetitive simultaneous estimates of CO obtained from the above devices were compared in 21 cardiac-surgery patients during the first 2 h post-surgery. Mean and absolute values for CO across the devices were compared by ANOVA, Bland-Altman, Pearson moment, and linear-regression analyses.
View Article and Find Full Text PDFActa Anaesthesiol Taiwan
June 2016
Department of Anesthesiology, National Taiwan University Hospital, Taipei, Taiwan. Electronic address:
Background: This study compared the cardiac output (CO) obtained from PiCCO with that obtained from the noninvasive NICOM method.
Methods: Twenty-one cirrhotic patients receiving liver transplantation were enrolled. During the operation, their CO was measured by the PiCCO system via the thermodilution method as the standard and by the NICOM method.
Objective: Cardiopulmonary monitoring (CPM) is rapidly progressing but data regarding CPM in Thai ICUs was unavailable. The objective of this study was to describe the situation, and gaps of CPM in Thai ICUs.
Material And Method: Data were retrieved from the ICU-RESOURCE I study database survey CPM was divided into two aspects of device and measurement methods.
Resuscitation
January 2012
Pediatric Intensive Care Department, Hospital General Universitario Gregorio Marañón, Universidad Complutense Madrid, Spain.
Aim: Bioreactance is a new non-invasive method for cardiac output measurement (NICOM). There are no studies that have analysed the utility of this technique in a pediatric animal model of hemorrhagic shock.
Methods: A prospective study was performed using 9 immature Maryland pigs weighing 9 to 12 kg was performed.