The new potent combination of antiretrovirals etravirine, darunavir, and ritonavir requires a new bioanalytical method for clinical pharmacology investigations and potential therapeutic drug monitoring. The development and validation of a novel LC-MS method for the simultaneous quantification of the most recently FDA-approved protease inhibitor and non-nucleoside reverse transcriptase inhibitor is described. This novel method was developed and validated using a sub-2 microm particle column, and provides excellent chromatographic separation and peak shape for all three analytes and internal standard. The method was validated over the range of 0.002-2.0 microg/mL. Intra- and inter-day accuracy of all analytes ranged from 88 to 106%, and intra- and inter-day precision was <7%. Dilution of samples 2-, 5-, and 10-fold maintained accuracy and precision, using a sample volume as low as 10 microL. Finally, the applicability of the method was investigated with clinical samples and external quality assurance proficiency testing samples.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2753357 | PMC |
http://dx.doi.org/10.1016/j.talanta.2009.06.005 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!