The new potent combination of antiretrovirals etravirine, darunavir, and ritonavir requires a new bioanalytical method for clinical pharmacology investigations and potential therapeutic drug monitoring. The development and validation of a novel LC-MS method for the simultaneous quantification of the most recently FDA-approved protease inhibitor and non-nucleoside reverse transcriptase inhibitor is described. This novel method was developed and validated using a sub-2 microm particle column, and provides excellent chromatographic separation and peak shape for all three analytes and internal standard. The method was validated over the range of 0.002-2.0 microg/mL. Intra- and inter-day accuracy of all analytes ranged from 88 to 106%, and intra- and inter-day precision was <7%. Dilution of samples 2-, 5-, and 10-fold maintained accuracy and precision, using a sample volume as low as 10 microL. Finally, the applicability of the method was investigated with clinical samples and external quality assurance proficiency testing samples.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2753357PMC
http://dx.doi.org/10.1016/j.talanta.2009.06.005DOI Listing

Publication Analysis

Top Keywords

method simultaneous
8
etravirine darunavir
8
darunavir ritonavir
8
intra- inter-day
8
method
5
novel lc-esi-ms
4
lc-esi-ms method
4
simultaneous determination
4
determination etravirine
4
ritonavir human
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!