Introduction: Cardiac resynchronization therapy (CRT) has the potential to improve the outcome of patients suffering from mechanical dyssynchrony and heart failure. It has been suggested that accurate quantification of baseline extent of mechanical dyssynchrony may lead to pre-selection of patients likely to respond to CRT. The standard deviation from a phase histogram (phaseSD), synchrony (S) and entropy (E) are parameters obtained from phase analysis of planar radionuclide angiography (RNA) that may provide an accurate means of assessing mechanical dyssynchrony. In this paper, the ability of phaseSD, S, and E to detect mechanical dyssynchrony was investigated and optimal values for image smoothing, histogram noise thresholding, and bin size were defined. Finally, the intra- and inter-observer reproducibility of the methodology was assessed.

Methods: PhaseSD, S, and E were calculated for 37 normal subjects (LVEF > 50%, end-diastolic volume < 120 mL, end-systolic volume < 60 mL, QRS < 120 ms, and normal wall motion) and 53 patients with mechanical dyssynchrony (LVEF < 30%, QRS > 120 ms, and typical LBBB). Receiver-operator characteristics (ROC) curves were created and the area under the curve (AUC), for each parameter, was determined using three different imaging filters (no filter and an order 5 Hann filter with cut-off of 5/50 and 10/50). The AUC was also determined using histogram threshold values varying between 0% and 50% (of the max amplitude value). Finally, AUC for E was determined for bins sizes varying between 1 degrees and 20 degrees . Inter- and intra-observer variability was calculated at optimal imaging values.

Results: No smoothing was found to maximize the AUC. The AUC was independent of histogram threshold value. However, a value of 20% provided optimal visualization of the phase image. The AUC was also independent of bin size. At the optimal imaging values, the sensitivity and specificity for all parameters for detection of mechanical dyssynchrony was measured to be 89-100%. Inter- and intra-observer correlation coefficients >0.99 were found for phaseSD, S and E.

Conclusions: Optimized planar RNA phase analysis parameters, phaseSD, S, and E, were able to detect mechanical dyssynchrony with low inter- and intra-observer variability. Studies assessing the ability of these parameters to predict CRT outcome are required.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s12350-009-9119-4DOI Listing

Publication Analysis

Top Keywords

mechanical dyssynchrony
32
phase analysis
12
inter- intra-observer
12
radionuclide angiography
8
analysis parameters
8
mechanical
8
dyssynchrony
8
phasesd detect
8
detect mechanical
8
bin size
8

Similar Publications

A leadless pacemaker (LP) is a modern alternative to a transvenous pacemaker, allowing certain complications to be avoided; however, some cannot be eliminated. To highlight the essential role of advanced speckle-tracking echocardiography (STE) in diagnosing pacing-induced cardiomyopathy (PICM) caused by an LP. A 79-year-old male, after LP implantation a year earlier, was admitted due to heart failure (HF).

View Article and Find Full Text PDF

Left bundle branch block-Innocent bystander, silent menace, or both.

Heart Rhythm

December 2024

Medical Faculty, Christian-Albrechts-University, Kiel, Germany; Life Sciences, University of Applied Sciences, Bremerhaven, Germany. Electronic address:

Left bundle branch block (LBBB) causes immediate electrical and mechanical dyssynchrony of the left ventricle (LV) and gradual structural damages in the Purkinje cells and myocardium. Mechanical dyssynchrony reduces the LV ejection fraction (EF) instantly, but only to ≈55% in an otherwise normal heart. Because of the heart's in-built functional redundancy, a patient with LBBB does not always notice the heart's reduced efficiency straightaway.

View Article and Find Full Text PDF

Echocardiography in the Assessment of Heart Failure Patients.

Diagnostics (Basel)

December 2024

Division of Cardiology, Department of Medical Biotechnologies, University of Siena, 53100 Siena, Italy.

Doppler echocardiography is the corner-stone of non-invasive investigation of patients with a clinical diagnosis of heart failure. It provides an accurate and quantitative assessment of cardiac structure and function. Furthermore, spectral Doppler measurement is an invaluable technique for estimating intracardiac pressures with their crucial value in the optimum management of heart failure patients, irrespective of ejection fraction.

View Article and Find Full Text PDF

Prognostic value of mechanical dyssynchrony in patients with heart failure: a systematic review.

BMC Cardiovasc Disord

November 2024

State Key Laboratory for Innovation and Transformation of Luobing Theory, Department of Cardiology, the First Affiliated Hospital of Nanjing Medical University, Nanjing, 210029, China.

Background: Heart failure (HF) significantly impacts quality of life and healthcare systems worldwide. Assessing left ventricular mechanical dyssynchrony (LVMD) is crucial for understanding cardiac function and optimizing treatments like cardiac resynchronization therapy (CRT). Phase analysis using gated single-photon emission computed tomography (SPECT) myocardial perfusion imaging (MPI) has shown promise in predicting outcomes, yet recent comprehensive reviews are lacking.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!