A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

A small tolerance for catheter displacement in high-dose rate prostate brachytherapy is necessary and feasible. | LitMetric

A small tolerance for catheter displacement in high-dose rate prostate brachytherapy is necessary and feasible.

Int J Radiat Oncol Biol Phys

Department of Radiation Oncology, Sir Charles Gairdner Hospital, Hospital Avenue, Nedlands, Perth, Western Australia 6009, Australia.

Published: March 2010

Purpose: We examined catheter displacement in patients treated with fractionated high-dose rate (HDR) brachytherapy boost for prostate cancer and the impact this had on tumor control probability (TCP). These data were used to make conclusions on an acceptable amount of displacement.

Methods And Materials: The last 20 patients treated with HDR brachytherapy boost for prostate cancer at our center in 2007 were replanned using simulated interstitial catheter displacements of 3, 6, 9, and 12 mm with originally planned dwell times. The computer-modeled dose-volume histograms for the clinical target volumes were exported and used to calculate the TCP of plans with displaced needles relative to the original plan. Actual catheter displacements were also measured before and after manual adjustment in all patients treated in 2007.

Results: In the 20 patients who were replanned for caudal catheter displacements of 3, 6, 9, and 12 mm, the median relative TCP was 0.998, 0.964, 0.797, and 0.265, respectively (p < 0.01 when all medians were compared). All patients replanned with a 3-mm displacement, compared with only 75% with a 6-mm displacement, had a relative TCP greater than 0.950. In the 91 patients treated in 2007, before adjustment, 82.3% of fractions had a displacement greater than 3 mm compared with 12.2% of fractions after adjustment.

Conclusions: Catheter displacement in HDR brachytherapy significantly compromises the TCP. The tolerance for these movements should be small (< or =3 mm). Correcting these displacements to within acceptable limits is feasible.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ijrobp.2009.03.052DOI Listing

Publication Analysis

Top Keywords

patients treated
16
catheter displacement
12
hdr brachytherapy
12
catheter displacements
12
high-dose rate
8
brachytherapy boost
8
boost prostate
8
prostate cancer
8
patients replanned
8
relative tcp
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!