Purpose: To evaluate retinal straylight and patient complaint scores 18 months after implantation with the AcrySof ReSTOR diffractive multifocal intraocular lens (IOL) (Alcon Laboratories Inc).

Methods: Retinal straylight was measured with the C-Quant (Oculus Optikgeräte GmbH) system in 40 eyes implanted with the AcrySof ReSTOR IOL (SA60D3) and in 40 eyes implanted with the AcrySof SA60AT monofocal IOL. Complaint scores were assessed by a questionnaire in both groups of patients at three different lighting conditions--day light, dimmed light, and at night.

Results: Seventy-five percent of eyes (30/40) of the ReSTOR group and 80% of eyes (32/40) of the SA60AT control group showed a retinal straylight value within or better than the normal range. No statistically significant differences between groups were found (P = .790). Patients of the ReSTOR group showed significantly more complaint scores at all assessed lighting conditions: increasing factors were 1.53, 2.02, and 2.66, for day light, dimmed light, and night, respectively (P < .0001). Comparing the amount of straylight, the multifocal group had 20% more straylight, albeit not significant at the 1% level.

Conclusions: The AcrySof ReSTOR IOL showed similar retinal straylight values compared to the monofocal IOL. Subjectively, patients with the diffractive IOL claimed significantly more glare for all light conditions, especially at night.

Download full-text PDF

Source
http://dx.doi.org/10.3928/1081597X-20090512-02DOI Listing

Publication Analysis

Top Keywords

retinal straylight
20
complaint scores
16
acrysof restor
12
scores months
8
months implantation
8
implantation acrysof
8
restor diffractive
8
eyes implanted
8
implanted acrysof
8
restor iol
8

Similar Publications

Purpose: Propose new terminology and evaluate the effectiveness of Therapeutic Refractive Vitrectomy (TRV) for selective removal of vitreous floaters and opacities (VFO) utilizing Standardized Kinetic Anatomical Functional Testing of VFO (SK VFO Test) and new ultra widefield (UWF) OCT imaging techniques.

Methods: Retrospective analysis. Twenty eyes underwent TRV for symptomatic VFO.

View Article and Find Full Text PDF

Purpose: To identify the biometric factors associated with postoperative visual performance after uneventful phacoemulsification with multifocal intraocular lens (MIOL) implantation.

Methods: In this retrospective cohort study, 72 eyes of 72 patients implanted with the HumanOptics Diff-aAY MIOL were included. Preoperative examination data including the white-to-white distance (WTW), anterior chamber depth (ACD), axial length and corneal astigmatism were gathered through the electronic medical records.

View Article and Find Full Text PDF

Purpose: Vision-degrading myodesopsia (VDM) from vitreous floaters significantly degrades vision and impacts visual quality of life (VQOL), but the relationship to light scattering is poorly understood. This study compared in vitro measures of light scatter and transmission in surgically excised human vitreous to preoperative indexes of vitreous structure, visual function, and VQOL.

Methods: Pure vitreous collected during vitrectomy from 8 patients with VDM had wide-angle straylight measurements and dark-field imaging, performed within 36 hours of vitrectomy.

View Article and Find Full Text PDF

Subjective Straylight Index: A Visual Test for Retinal Contrast Assessment as a Function of Veiling Glare.

J Imaging

April 2024

Departamento de Física Aplicada, Facultad de Ciencias, Universidad de Zaragoza, 50009 Zaragoza, Spain.

Spatial aspects of visual performance are usually evaluated through visual acuity charts and contrast sensitivity (CS) tests. CS tests are generated by vanishing the contrast level of the visual charts. However, the quality of retinal images can be affected by both ocular aberrations and scattering effects and none of those factors are incorporated as parameters in visual tests in clinical practice.

View Article and Find Full Text PDF

Alcohol Effects on Drivers' Speed Management: The Influence of Visual Performance and Road Complexity.

J Stud Alcohol Drugs

November 2024

Laboratorio de Ciencias de la Visión y Aplicaciones, Departamento de Óptica, Facultad de Ciencias de la Universidad de Granada, Fuente Nueva s/n 18071 Granada, Spain.

Objective: The aim of this work is to assess how drivers adapt their driving speed as a self-regulation mechanism when driving under the influence of alcohol, and the influence of alcohol consumption, visual performance, road complexity, and personal traits.

Method: Thirty-one volunteers took part in the study. All of them underwent three experimental sessions: baseline (no alcohol), Alcohol 1 (low-moderate dose), and Alcohol 2 (moderate-high dose).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!