Addressing a systematic vibration artifact in diffusion-weighted MRI.

Hum Brain Mapp

Centre for Functional Magnetic Resonance Imaging of the Brain, John Radcliffe Hospital, University of Oxford, Oxford, United Kingdom.

Published: February 2010

We have identified and studied a pronounced artifact in diffusion-weighted MRI on a clinical system. The artifact results from vibrations of the patient table due to low-frequency mechanical resonances of the system which are stimulated by the low-frequency gradient switching associated with the diffusion-weighting. The artifact manifests as localized signal-loss in images acquired with partial Fourier coverage when there is a strong component of the diffusion-gradient vector in the left-right direction. This signal loss is caused by local phase ramps in the image domain which shift the apparent k-space center for a particular voxel outside the covered region. The local signal loss masquerades as signal attenuation due to diffusion, severely disrupting the quantitative measures associated with diffusion-tensor imaging (DTI). We suggest a way to improve the interpretation of affected DTI data by including a co-regressor which accounts for the empirical response of regions affected by the artifact. We also demonstrate that the artifact may be avoided by acquiring full k-space data, and that subsequent increases in TE can be avoided by employing parallel acceleration.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6870579PMC
http://dx.doi.org/10.1002/hbm.20856DOI Listing

Publication Analysis

Top Keywords

artifact diffusion-weighted
8
diffusion-weighted mri
8
signal loss
8
artifact
6
addressing systematic
4
systematic vibration
4
vibration artifact
4
mri identified
4
identified studied
4
studied pronounced
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!