Introduction: During the preparation of curved canals, rotary nickel-titanium (NiTi) instruments are subjected to cyclic fatigue, which can lead to instrument fracture. Although several factors may influence the cyclic fatigue resistance of instruments, the role of the rotational speed remains uncertain. This study was intended to evaluate the effects of rotational speed on the number of cycles to fracture of rotary NiTi instruments.

Methods: ProTaper Universal instruments F3 and F4 (Maillefer SA, Ballaigues, Switzerland) were used in an artificial curved canal under rotational speeds of 300 rpm or 600 rpm. The artificial canal was made of stainless steel, with an inner diameter of 1.5 mm, total length of 20 mm, and arc at the end with a curvature radius of 6 mm. The arc length was 9.4 mm and 10.6 mm on the straight part. The number of cycles required to fracture was recorded. Fractured surfaces and the helical shafts of the fractured instruments were analyzed by scanning electron microscopy.

Results: The results showed approximately a 30% reduction in the observed number of cycles to fracture as rotational speed was increased from 300 to 600 RPM (p < 0.05). The morphology of the fractured surface was always of ductile type, and no plastic deformation was observed on the helical shaft of fractured instruments.

Conclusions: The present findings for both F3 and F4 ProTaper instruments revealed that the increase in rotational speed significantly reduced the number of cycles to fracture.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.joen.2009.04.003DOI Listing

Publication Analysis

Top Keywords

rotational speed
20
number cycles
16
cyclic fatigue
12
cycles fracture
12
rotary nickel-titanium
8
600 rpm
8
instruments
6
speed
5
fracture
5
rotational
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!