Detection of molecules with wide unresolved rotationa-lvibrational absorption bands is demonstrated by using Quartz Enhanced Photoacoustic Spectroscopy and an amplitude modulated, high power, thermoelectrically cooled quantum cascade laser operating at 8.4 mum in an external cavity configuration. The laser source exhibits single frequency tuning of 135 cm-1 with a maximum optical output power of 50 mW. For trace-gas detection of Freon 125 (pentafluoroethane) at 1208.62 cm-1 a normalized noise equivalent absorption coefficient of NNEA=2.64x10(-9) cm?(-1)W/Hz(1/2)was obtained. Noise equivalent sensitivity at ppbv level as well as spectroscopic chemical analysis of a mixture of two broadband absorbers (Freon 125 and acetone) with overlapping absorption spectra were demonstrated.