Spectrally-encoded miniature endoscopy uses a single optical fiber and wavelength division multiplexing to obtain macroscopic images through miniature, flexible probes. In turn, it has the potential to enable two- and three-dimensional imaging within the body at locations that are currently difficult to access with conventional endoscopes. Here we present a novel detection scheme for spectrally-encoded endoscopy using spectral-domain interferometry. Compared to previous time-domain configurations, this new detection method results in greater than 1000-fold increase in sensitivity (77 dB), a 6-fold increase in imaging speed (30 volumes per second), and a 2-fold increase in depth range (2.8 mm). We demonstrate spectrally-encoded, spectral-domain detection by conducting video-rate, three-dimensional imaging in a variety of specimens, including the paws of a mouse embryo and excised human ear bones. Our results show that this new technology enables video rate spectrally-encoded endoscopy and will therefore be useful for a variety of minimally invasive medical applications.

Download full-text PDF

Source
http://dx.doi.org/10.1364/oe.15.002432DOI Listing

Publication Analysis

Top Keywords

spectrally-encoded endoscopy
12
three-dimensional imaging
8
spectral-domain spectrally-encoded
4
endoscopy
4
spectrally-encoded
4
endoscopy spectrally-encoded
4
spectrally-encoded miniature
4
miniature endoscopy
4
endoscopy single
4
single optical
4

Similar Publications

Introduction: The acoustic reflex is the active response of the middle ear to loud sounds, altering the mechanical transfer function of the acoustic energy into the inner ear. Our goal was to observe the effect of the acoustic reflex on the tympanic membrane by identifying a significant nonlinear increase in membrane oscillations.

Methods: By using interferometric spectrally encoded endoscopy, we record the membrane oscillations over time in response to a loud, 200-ms-long acoustic stimulus.

View Article and Find Full Text PDF

The wide frequency range of the human hearing could be narrowed by various pathologies in the middle ear and in the tympanic membrane that lead to conductive hearing loss. Diagnosing such hearing problems is challenging, however, often relying on subjective hearing tests supported by functional tympanometry. Here we present a method for in vivo 2D mapping of the impulse response of the tympanic membrane, and demonstrate its potential on a healthy human volunteer.

View Article and Find Full Text PDF

RGB-color forward-viewing spectrally encoded endoscope using three orders of diffraction.

Biomed Opt Express

February 2021

Healthcare Optics Research Laboratory, Canon U.S.A., Inc., Cambridge, MA 02139, USA.

Spectrally encoded endoscopy (SEE) is an ultra-miniature endoscopy technology that encodes each spatial location on the sample with a different wavelength. One challenge in SEE is achieving color imaging with a small probe. We present a novel SEE probe that is capable of conducting real-time RGB imaging using three diffraction orders (6th order diffraction of the blue spectrum, 5th of green, and 4th of red).

View Article and Find Full Text PDF

Characterization of oviduct ciliary beat frequency using real time phase resolved Doppler spectrally encoded interferometric microscopy.

Biomed Opt Express

November 2019

Beckman Laser Institute, Department of Biomedical Engineering, University of California, Irvine, 1002 Health Sciences Road East, Irvine, CA 92612, USA.

Ciliary activity, characterized by the coordinated beating of ciliary cells, generates the primary driving force for oviduct tubal transport, which is an essential physiological process for successful pregnancies. Malfunction of the cilium in the fallopian tube, or oviduct, may increase the risk of infertility and tubal pregnancy that can result in maternal death. While many ex-vivo studies have been carried out using bright field microscopy, this technique is not feasible for the in-vivo investigation of oviduct ciliary beating frequency (CBF).

View Article and Find Full Text PDF

Background And Objective: Spectrally encoded endoscopy (SEE) is an optical imaging technology that uses spatial wavelength multiplexing to conduct endoscopy in miniature, small diameter probes. Contrary to the previous side-viewing SEE devices, forward-viewing SEE probes are advantageous as they provide a look ahead that facilitates navigation and surveillance. The objective of this work was to develop a miniature forward-viewing SEE probe with a wide field of view and a high spatial resolution.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!