Unlabelled: (82)Rb cardiac PET allows the assessment of myocardial perfusion with a column generator in clinics that lack a cyclotron. There is evidence that the quantitation of myocardial blood flow (MBF) and coronary flow reserve (CFR) with dynamic (82)Rb PET is feasible. The objectives of this study were to determine the accuracy and reproducibility of MBF estimates from dynamic (82)Rb PET by using our methodology for generalized factor analysis (generalized factor analysis of dynamic sequences [GFADS]) and compartment analysis.

Methods: Reproducibility was evaluated in 22 subjects undergoing dynamic rest and dipyridamole stress (82)Rb PET studies at a 2-wk interval. The inter- and intraobserver variability of MBF quantitation with dynamic (82)Rb PET was assessed with 4 repeated estimations by each of 4 observers. Accuracy was evaluated in 20 subjects undergoing dynamic rest and dipyridamole stress PET studies with (82)Rb and (13)N-ammonia, respectively. The left ventricular and right ventricular blood pool and left ventricular tissue time-activity curves were estimated by GFADS. MBF was estimated by fitting the blood pool and tissue time-activity curves to a 2-compartment kinetic model for (82)Rb and to a 3-compartment model for (13)N-ammonia. CFR was estimated as the ratio of peak MBF to baseline MBF.

Results: The reproducibility of the MBF estimates in repeated (82)Rb studies was very good at rest and during peak stress (R(2)= 0.935), as was the reproducibility of the CFR estimates (R(2) = 0.841). The slope of the correlation line was very close to one for the estimation of MBF (0.986) and CFR (0.960) in repeated (82)Rb studies. The intraobserver reliability was less than 3% for the estimation of MBF at rest and during peak stress as well as for the estimation of CFR. The interobserver reliabilities were 0.950 at rest and 0.975 at peak stress. The correlation between myocardial flow estimates obtained at rest and those obtained during peak stress in (82)Rb and (13)N-ammonia studies was very good (R(2) = 0.857). Bland-Altman plots comparing CFR estimated with (82)Rb and CFR estimated with (13)N-ammonia revealed an underestimation of CFR with (82)Rb compared with (13)N-ammonia; the underestimation was within +/-1.96 SD.

Conclusion: MBF quantitation with GFADS and dynamic (82)Rb PET demonstrated excellent reproducibility as well as intra- and interobserver reliability. The accuracy of the absolute quantitation of MBF with factor and compartment analyses and dynamic (82)Rb PET was very good, compared with that achieved with (13)N-ammonia, for MBF of up to 2.5 mL/g/min.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3133618PMC
http://dx.doi.org/10.2967/jnumed.104.007831DOI Listing

Publication Analysis

Top Keywords

82rb pet
28
dynamic 82rb
20
peak stress
16
82rb
15
cfr estimated
12
rest peak
12
mbf
11
pet
10
myocardial blood
8
blood flow
8

Similar Publications

Background: The majority of functional ischemia tests in patients with suspected chronic coronary syndromes (CCS) yield normal results. Implementing gatekeepers for patient preselection, such as pretest probability (PTP) and/or coronary artery calcium score (CACS), could reduce the number of normal scan results, radiation exposure and costs. However, the efficacy and safety of these approaches remain unclear.

View Article and Find Full Text PDF

Importance: Long-term use of anabolic androgenic steroids (AASs) is associated with a high risk of left ventricular hypertrophy, heart failure with reduced systolic function, and early sudden death, with the mechanism of progression being understudied. Early and persistent impaired myocardial microcirculation could be of clinical importance and a potential underlying mechanism of frequent and early cardiac disease among individuals with AAS use and a future potential target for intervention.

Objective: To investigate coronary microcirculation by measuring myocardial flow reserve (MFR) in men with current and former AAS use compared with controls with no prior AAS use, using cardiac rubidium 82 (82Rb) positron emission tomography/computed tomography (PET/CT).

View Article and Find Full Text PDF

The impact of long axial field of view (LAFOV) PET on oncologic imaging.

Eur J Radiol

December 2024

King's College London & Guy's and St Thomas' PET Centre, School of Biomedical Engineering and Imaging Sciences, King's College, London SE1 7EH, UK. Electronic address:

The development of long axial field of view (LAFOV) positron emission tomography coupled with computed tomography (PET/CT) scanners might be considered the biggest step forward in PET imaging since it became a mainstream clinical modality. Despite increased capital and maintenance costs and data storage requirements, the improvement in image quality, significantly faster acquisition times and lower radiopharmaceutical administered activities, allow a high quality and more efficient clinical service. This step change in technology overcomes some of the limitations of standard short axial field of view scanners.

View Article and Find Full Text PDF

Noise-aware dynamic image denoising and positron range correction for Rubidium-82 cardiac PET imaging via self-supervision.

Med Image Anal

February 2025

Department of Biomedical Engineering, Yale University, USA; Department of Radiology and Biomedical Imaging, Yale University, USA. Electronic address:

Rubidium-82 (Rb) is a radioactive isotope widely used for cardiac PET imaging. Despite numerous benefits of Rb, there are several factors that limits its image quality and quantitative accuracy. First, the short half-life of Rb results in noisy dynamic frames.

View Article and Find Full Text PDF

Background: Transient ischemic dilation (TID) of the left ventricular (LV) cavity is considered a high-risk marker in patients with abnormal single photon emission computed tomography (SPECT) myocardial perfusion imaging (MPI). Stress image acquisition with rubidium-82 (Rb) PET occurs at peak stress compared to 30-60 minutes post-stress with SPECT. We aimed to evaluate the prognostic value of TID in patients undergoing Rb PET MPI.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!