Accuracy of whole-body plethysmography requires biological calibration.

Chest

Copenhagen Prospective Study on Asthma in Children, Danish Pediatric Asthma Center, Health Sciences, University of Copenhagen, Copenhagen University Hospital, Gentofte, Denmark. Electronic address:

Published: June 2009

Background: Specific airway resistance (sRaw) measured by whole-body plethysmography in young children is increasingly used in research and clinical practice. The method is precise and feasible. However, there is no available method for calibration of the resistance measure, which raises concern of accuracy. Our aim was to determine the agreement of sRaw measurements in six centers and expand normative sRaw values for nonasthmatic children including these centers.

Method: Identical hardware with different software versions was used at the six centers. Measurements followed a standard operating procedure: (1) seven healthy young children were brought to each of the six centers for sRaw measurements; and (2) 105 healthy preschool children (52 boys; mean age, 5.1 years; interquartile range, 4.3 to 6.0) were recruited locally for sRaw measurements.

Results: (1) The sRaw of the seven-children study group was significantly lower at two centers compared with the other four centers, and one center had significantly higher sRaw than all the other centers (p < 0.05). Error in the factory settings of the software was subsequently discovered in one of the deviating centers. (2) Normative data (105 preschool children) were generated and were without significant difference between centers and independent of height, weight, age, and gender. We subsequently pooled these normative data (105 children) with our previous data from 121 healthy young children (overall mean sRaw, 1.27; SD, 0.25).

Conclusion: Control using biological standards revealed errors in the factory setting and highlights the need for developing methods for verification of resistance measures to assure accuracy. Normative data were subsequently generated. Importantly, other centers using such normative data should first consider proper calibration before applying reference values.

Download full-text PDF

Source
http://dx.doi.org/10.1378/chest.08-1555DOI Listing

Publication Analysis

Top Keywords

normative data
16
young children
12
centers
9
whole-body plethysmography
8
sraw
8
sraw measurements
8
healthy young
8
preschool children
8
centers normative
8
data 105
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!