This paper describes a simple method for the simultaneous determination and confirmation of the indandione rodenticides in serum. After samples were extracted with 10% (v/v) methanol in acetonitrile and cleaned by solid-phase extraction, chromatographic separation was performed on an IonPac AS11 analytical column (250 x 4.0 mm) using gradient KOH eluent with 10% (v/v) methanol as organic modifier. Confirmation was depended on the extensive fragmentation of the indandione molecule under MS/MS conditions which provides sufficient structural information. Quantification was performed by negative electrospray ionization in multiple reaction monitoring mode. All the method parameters were validated. It was confirmed that this method could be used in clinical diagnosis and forensic toxicology.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/bmc.1246 | DOI Listing |