Background: Total hip arthroplasty without cement is frequently performed in young active patients, but only limited outcomes data are available after durations of follow-up of more than fifteen years.

Methods: We retrospectively evaluated the clinical and radiographic results of a consecutive series of 154 total hip arthroplasties (in 141 patients) performed with an uncemented grit-blasted straight tapered titanium femoral stem combined with a threaded socket in patients under the age of fifty-five years. The median duration of follow-up was seventeen years. Clinical results were evaluated with use of the Harris hip score. The canal fill index was used as the criterion to determine the adequacy of stem sizing. Kaplan-Meier survivorship analysis was performed to predict long-term outcomes.

Results: The stem was undersized, with a canal fill index of < or =80%, in forty-one hips (27%). Late aseptic loosening of the stem occurred in four femora, and the femoral component was undersized in all four. These four stems were stable for ten years and then underwent progressive subsidence, which was associated with pain. Five stems were revised because of a late postoperative periprosthetic fracture following trauma. Localized proximal femoral osteolysis was seen in seven hips without signs of loosening. Survivorship of the stem with revision for any reason as the end point was estimated to be 90% (95% confidence interval, 87% to 97%) at twenty years. Survivorship with aseptic loosening as the end point was estimated to be 95% (95% confidence interval, 91% to 99%) at twenty years. Sixty-seven (44%) of the threaded uncemented acetabular components were revised during the follow-up period.

Conclusions: After a minimum duration of follow-up of fifteen years, the survival of this type of femoral component is excellent in individuals younger than fifty-five years. The main mode of stem failure was a periprosthetic fracture due to trauma, or late aseptic loosening in a small percentage of the hips in which the femoral implant was undersized. The high rate of failure of the acetabular components was attributable to a poor design that is no longer in use.

Download full-text PDF

Source
http://dx.doi.org/10.2106/JBJS.H.00297DOI Listing

Publication Analysis

Top Keywords

fifty-five years
12
aseptic loosening
12
uncemented grit-blasted
8
grit-blasted straight
8
straight tapered
8
tapered titanium
8
younger fifty-five
8
years
8
total hip
8
follow-up fifteen
8

Similar Publications

Teaching high quality paediatric basic life support to laypeople: The development and evaluation of a virtual simulation game.

Resusc Plus

January 2025

Department of Paediatrics, Division of Paediatric Critical Care, CHEO, 401 Smyth Rd, Ottawa, Ontario K1H 8L1, Canada.

Background: Self-directed training has been recognized as a reasonable alternative to traditional instructor-led formats to teach laypeople Basic Life Support (BLS). Virtual tools can facilitate high-quality self-directed resuscitation education; however, their role in teaching paediatric BLS remains unclear due to limited empiric evaluation and suboptimal design of existing tools.

Aim: We describe the development and evaluation of a virtual simulation game (VSG) designed to teach high-quality paediatric BLS using a self-directed, online format with integrated deliberate practice and feedback.

View Article and Find Full Text PDF

Background: Electroconvulsive therapy (ECT) is a highly effective treatment for schizophrenia and mood disorders; however, most evidence is derived from the adult population, with less evidence in adolescents. We sought to determine the use of ECT in adolescents in the Institute of Mental Health (IMH) and evaluate the treatment outcome.

Methods: We conducted a retrospective naturalistic analysis of ECT registry data of patients aged from 10 to 19 years from March 2017 to March 2023.

View Article and Find Full Text PDF

Background: The expanded polytetrafluoroethylene (ePTFE) valved conduit (VC) has been reported for pulmonary valve replacement (PVR). The purpose of this study was to review long-term outcomes of our trileaflet ePTFE VC.

Methods: This multicenter study was performed with institutional review board approval from each institution.

View Article and Find Full Text PDF

Background: Detection of alloantibodies associated with the JK system may be tricky. They are nevertheless associated with transfusion reactions and their detection is crucial.

Study Design And Methods: Retrospectively and over a period of 7 years, we compared the results obtained using two different assays for antibody detection.

View Article and Find Full Text PDF

This study aimed to analyze the relationship between cutaneous microcirculation reactivity, retinal circulation, macrocirculation function, and specific adhesion molecules in young patients with uncomplicated type 1 diabetes. Fifty-five patients with type 1 diabetes mellitus (T1DM), aged 8 to 18 years, were divided into subgroups based on skin microcirculation reactivity. The cutaneous microcirculatory vessels were considered reactive if post-test PORH coverage increased compared to pre-test coverage.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!