Background: Qualitative point-of-care (POC) tests for human chorionic gonadotropin (hCG) vary in their ability to detect purified hCG variants and there is data to suggest that over-the-counter (OTC) devices might also display similar variability. This could potentially influence the detection of urine hCG in early pregnancy.

Methods: Six OTC devices were tested for their ability to detect 5 hCG variants. Ten early pregnancy urine specimens were selected for their diverse expression of hCG variants. The samples were tested with 6 brands of POC and 6 OTC devices.

Results: OTC devices consistently recognized intact hCG, hCGn, and hCGbeta. hCGbetan was consistently recognized by 4 out of 6 brands. One brand inconsistently recognized hCGbetacf. OTC and POC devices varied greatly in their ability to detect hCG in early pregnancy urine, despite the fact that urine samples were adjusted to the same intact hCG concentration. Interestingly, we found that the OTC devices had better analytical sensitivity than the POC devices. Clinitest and First Response demonstrated the lowest hCG detection limits for POC and OTC devices, respectively.

Conclusions: Both OTC and POC devices are capable of detecting hCG concentrations in early pregnancy urine, and OTC devices demonstrated better analytical sensitivity relative to POC devices.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.cca.2009.05.018DOI Listing

Publication Analysis

Top Keywords

otc devices
24
hcg variants
16
early pregnancy
16
poc devices
16
hcg
12
detect hcg
12
ability detect
12
pregnancy urine
12
devices
11
otc
9

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!