Objective: To assess the repeatability and reproducibility of central corneal thickness (CCT) measurement by the commercial available ultrasonic pachymeter in Songklanagarind Hospital.

Material And Method: Seventy eyes underwent two sessions of CCT measurements, each session consisting of two CCT measurements, performed by two different operators. The mean and standard deviation of the mean differences were calculated. The interoperator repeatability and interoperator reproducibility were calculated by means of intraclass correlation coefficient (ICC). Agreement was analyzed by means of Bland-Altman plots.

Results: The mean CCT for 70 eyes was 537.61 +/- 26.66 mm. The intraoperator repeatability study of the first operator, the mean difference between the repeated measurements was 0.49 mm. The ICC was 0.985 (p < 0.001). For the second operator, the mean difference between the repeated measurements was 0.57 mm. The ICC was 0.935 (p < 0.001). A Bland-Altman plot of both operators showed narrow limits of agreement with respect to CCT for both operators. The mean measurement difference between operator was 0.96 mm and the ICC for reproducibility study was 0.979 (p < 0.01). The Bland-Altman plot showed narrow limits of agreement with respect to CCT.

Conclusion: The measurement of CCT using the ultrasonic pachymeter is highly repeatable and reproducible.

Download full-text PDF

Source

Publication Analysis

Top Keywords

intraoperator repeatability
8
repeatability interoperator
8
interoperator reproducibility
8
central corneal
8
corneal thickness
8
ultrasonic pachymeter
8
cct measurements
8
operator difference
8
difference repeated
8
repeated measurements
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!