Single-layer planar on-chip flow cytometer using microfluidic drifting based three-dimensional (3D) hydrodynamic focusing.

Lab Chip

Department of Engineering Science and Mechanics, The Pennsylvania State University, University Park, PA 16802, USA.

Published: June 2009

In this work, we demonstrate an on-chip microfluidic flow cytometry system based on a three-dimensional (3D) hydrodynamic focusing technique, microfluidic drifting. By inducing Dean flow in a curved microfluidic channel, microfluidic drifting can be used to hydrodynamically focus cells or particles in the vertical direction and enables the 3D hydrodynamic focusing in a single-layer planar microfluidic device. Through theoretical calculation, numerical simulation, and experimental characterization, we found that the microfluidic drifting technique can be effectively applied to three-dimensionally focus microparticles with density and size equivalent to those of human CD4+ T lymphocytes. In addition, we developed a flow cytometry platform by integrating the 3D focusing device with a laser-induced fluorescence (LIF) detection system. The system was shown to provide effective high-throughput flow cytometry measurements at a rate of greater than 1700 cells s(-1).

Download full-text PDF

Source
http://dx.doi.org/10.1039/b820138bDOI Listing

Publication Analysis

Top Keywords

microfluidic drifting
16
hydrodynamic focusing
12
flow cytometry
12
single-layer planar
8
based three-dimensional
8
three-dimensional hydrodynamic
8
microfluidic
7
flow
5
planar on-chip
4
on-chip flow
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!