Luminescence detection has great potential for use in small volume assays for high-throughput applications. However, the need for automated injection of detection reagents dictates an open well format, making small volume luminescence assays very susceptible to sample evaporation effects. Here we describe a technique for reducing evaporation of small sample volumes that uses layering of silicone oil on solution surfaces but still allows the use of automated injection.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ab.2009.01.034 | DOI Listing |