Background: Several studies have shown video laryngoscopy to be a useful technique in the management of patients in whom glottic exposure by direct laryngoscopy is difficult. We conducted this study as a preliminary investigation comparing the Storz DCI Miller 1 video laryngoscope (VL, Karl Storz GmbH, Tuttlingen, Germany) and direct laryngoscopy with a Miller 1 laryngoscope (DL) in an infant manikin model simulating difficult direct laryngoscopy. We hypothesized that compared with DL, VL would provide a better glottic view but would be associated with a longer time to intubation because of the different skill set required when using video intubation.

Methods: A Laerdal infant airway management training manikin (Laerdal Medical, Wappingers Falls, NY) was adapted using cloth tape to limit cervical spine mobility. Thirty-two attending pediatric anesthesiologists attempted tracheal intubation of the infant manikin using VL and DL in randomized order. The best laryngeal view with each laryngoscope and time to intubation were documented.

Results: There was a significant difference in the distributions of laryngoscopy grades between VL and DL (P < 0.001), with the VL giving a better laryngeal view. Forty percent of anesthesiologists reported a Grade 3 or 4 view with DL; all of which were converted to Grades 1 and 2 with VL. The median grade with interquartile range was two (2-3) for DL and one (1-2) for VL (P < 0.001). Seventy-eight percent of participants reported an improvement of at least one grade in laryngeal view with VL compared with DL. There were two failed intubations using DL and none using VL. Time to intubation was similar between the two techniques.

Conclusions: The Storz Miller 1 VL blade improved glottic exposure in a simulated difficult laryngoscopy compared with direct laryngoscopy with a standard Miller 1 blade without increasing the time to intubation.

Download full-text PDF

Source
http://dx.doi.org/10.1213/ane.0b013e3181a1a600DOI Listing

Publication Analysis

Top Keywords

direct laryngoscopy
16
time intubation
16
laryngeal view
12
storz miller
8
miller video
8
video laryngoscope
8
glottic exposure
8
infant manikin
8
miller blade
8
laryngoscopy
7

Similar Publications

Objectives: Injection laryngoplasty (IL) is commonly performed for glottic insufficiency and has historically been performed under general anesthesia via direct laryngoscopy (DL), with an increasing number of procedures being performed in the office setting via flexible laryngoscopy (FL). Existing literature regarding complications of IL primarily addresses immediate peri-procedural complications and adverse reactions to the injection material. This is the first study utilizing a large multi-institutional database comparing complications of IL performed via DL versus FL.

View Article and Find Full Text PDF

Background: The Neonatal Resuscitation Program recommends direct laryngoscopy (DL) as the primary method for neonatal intubation. Video laryngoscopy (VL) is suggested as an option, particularly for training novice operators or for intubating infants with difficult airways. The programme outlines specific steps for intubation, including managing the external environment and techniques for visualising key anatomical landmarks.

View Article and Find Full Text PDF

Background: Airway management is a critical component of the care of patients experiencing cardiac arrest, but data from randomized trials on the use of video vs direct laryngoscopy for intubation in the setting of cardiac arrest are limited. Current AHA guidelines recommend placement of an endotracheal tube either during CPR or shortly after return of spontaneous circulation but do not provide guidance around intubation methods, including the choice of laryngoscope.

Research Question: Does use of video laryngoscopy improve the incidence of successful intubation on the first attempt, compared to use of direct laryngoscopy, among adults undergoing tracheal intubation after experiencing cardiac arrest?

Study Design And Methods: This secondary analysis of the Direct versus Video Laryngoscope (DEVICE) trial compared video laryngoscopy versus direct laryngoscopy in the subgroup of patients who were intubated following cardiac arrest.

View Article and Find Full Text PDF

Background: Proficiency in endotracheal intubation (ETI) is essential for medical professionals and its training should start at medical schools; however, large caseload may be required before achieving an acceptable success rate with direct laryngoscopy. Video laryngoscopy has proven to be an easier alternative for intubation with a faster learning curve, but its availability in medical training may be an issue due to its high market prices. We devised a low-cost 3-dimensionally printed video laryngoscope (3DVL) and performed a randomized trial to evaluate if the intubation success rate on the first attempt with this device is noninferior to a standard commercially available video laryngoscope (STVL).

View Article and Find Full Text PDF

Background: To evaluate the role of ultrasound (US) in the assessment of the airway and to determine whether US has the potential to serve as effective, noninvasive and less time-consuming method for the diagnosis of difficult intubation in ICU patients.

Patients And Methods: This cross-sectional study was carried in 152 critically ill patients who underwent intubation in the ICU from December 2022 to April 2024. Prior to intubation thyromental height (TMH) and hyomental distance ratio (HMD-R) was measured using a scale and distance from skin to hyoid bone (SHB) and distance from skin to thyrohyoid membrane (STM) was measured using a US.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!