A simple CE method has been developed for the separation and determination of thyroxine (T4) enantiomers in pharmaceutical formulations. The method was based on ligand-exchange mechanism using a Cu(II)/L-proline complex as chiral selector. The effects of different parameters affecting separation such as chiral selector concentration, organic additive, buffer pH and temperature were investigated. A baseline separation of the two enantiomers was obtained at a Cu(II)/L-proline ratio of 1:8 in a borate buffer (15 mmol/L, pH 9.6) containing 10% v/v acetonitrile. Under the optimized conditions, precision linearity range and detection limits of the developed enantioselective CE method were evaluated and compared using two different detection systems: conventional UV detection at 226 nm and iodine (127I)specific detection ("chiral speciation") with ICP-MS. Both methodologies show adequate analytical performance characteristics with detection limits around 0.30 microg/mL for each enantiomer of T4. Finally, a levothroid pharmaceutical formulation sample was successfully analyzed using both developed methods CE-UV and CE-ICP-MS.

Download full-text PDF

Source
http://dx.doi.org/10.1002/elps.200800731DOI Listing

Publication Analysis

Top Keywords

determination thyroxine
8
thyroxine enantiomers
8
chiral selector
8
detection limits
8
detection
6
enantioselective determination
4
enantiomers ligand-exchange
4
ligand-exchange absorbance
4
absorbance icp-ms
4
icp-ms detection
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!