Study Objective: To compare efficacy of rollerball endometrial ablation with low-voltage (cut) versus high-voltage (coag) waveforms.

Design: Pilot comparative clinical study (Canadian Task Force Classification II-1).

Setting: University-affiliated teaching hospital.

Patients: Fifty premenopausal women with menorrhagia.

Intervention: Women with menorrhagia were allocated randomly to thermal destruction of the endometrium by a 5-mm rollerball with unmodulated cutting current or modulated coagulating current. Complication rate, clinical outcomes, and need for reintervention were evaluated.

Measurements And Main Results: At 2 years of follow-up, the reintervention rate was 26.3% in the cutting waveform group versus 31.4% in the coagulating waveform group. This difference was not statistically significant. Hysterectomy was performed in 3 (14%) women in the cutting waveform group and 5 (20%) women in the coagulating waveform group. There were no complications in either group.

Conclusion: Both cutting and coagulating waveforms are equally effective for hysteroscopic endometrial ablation with the rollerball.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jmig.2009.03.001DOI Listing

Publication Analysis

Top Keywords

waveform group
16
endometrial ablation
12
clinical outcomes
8
low-voltage cut
8
cut versus
8
versus high-voltage
8
high-voltage coag
8
hysteroscopic endometrial
8
ablation rollerball
8
cutting waveform
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?