Background: Many studies have indicated that acceleromyography and mechanomyography cannot be used interchangeably. To improve the agreement between the two methods, it has been suggested to use a preload and to refer all train-of-four (TOF) ratios to the control TOF (normalization) when using acceleromyography. The first purpose of this study was to test whether a preload applied to acceleromyography would increase the precision and the agreement with mechanomyography. The second purpose was to evaluate whether normalization would improve the agreement with mechanomyography.

Methods: Sixty patients were randomized to acceleromyography with or without a preload (Hand Adapter; Organon, Oss, the Netherlands). On the contralateral arm, mechanomyography was used. Anesthesia was induced with propofol and an opioid, and neuromuscular block with 0.6 mg/kg rocuronium. The precision and the bias and limits of agreement (with or without normalization) between the methods were evaluated using TOF stimulation.

Results: Preload improved the precision of acceleromyography by 21%, but it also increased the mean control TOF ratio from 1.07 to 1.13. Normalization of the acceleromyographic TOF ratios diminished the bias to mechanomyography during recovery (e.g., from 0.15 to 0.05 at TOF 0.90); when the mechanomyographic TOF values were normalized as well, the bias was eliminated. However, normalization did not exclude wide individual differences between acceleromyography and mechanomyography (+/- 0.10-0.20 at TOF 0.90).

Conclusion: Preload increases the precision of acceleromyography, and normalization of the TOF ratios decreases bias in relation to mechanomyography. When both acceleromyography and mechanomyography are normalized, there is no significant bias between the two methods.

Download full-text PDF

Source
http://dx.doi.org/10.1097/ALN.0b013e3181a4f239DOI Listing

Publication Analysis

Top Keywords

acceleromyography mechanomyography
12
tof ratios
12
tof
9
mechanomyography
8
acceleromyography
8
improve agreement
8
control tof
8
precision acceleromyography
8
preload
6
normalization
6

Similar Publications

Background: The accuracy and precision of currently available, widely used acceleromyograph and electromyograph neuromuscular blockade monitors have not been well studied. In addition, the normalization of the train-of-four ratio from acceleromyography (train-of-four ratio [T4/T1] divided by the baseline train-of-four ratio) has not been validated in comparison to mechanomyography.

Methods: Enrolled patients had surgery under general anesthesia with a supraglottic airway and without any neuromuscular blocking drugs.

View Article and Find Full Text PDF

The administration of general anesthesia is a crucial aspect of surgery. However, it can pose significant risks to patients, such as respiratory depression and prolonged neuromuscular blockade. To avoid such complications, it is essential to monitor neuromuscular transmission during anesthesia.

View Article and Find Full Text PDF

Evaluation of a three-axial acceleromyography monitor in dogs compared with mechanomyography.

Vet Anaesth Analg

September 2023

Department of Clinical Sciences, College of Veterinary Medicine, Cornell University, Ithaca, NY, USA.

Objective: To compare the ratio of the train-of-four (TOF) and double burst stimulation (DBS) obtained with three-axial acceleromyography (AMG) and mechanomyography (MMG) in dogs during recovery from a rocuronium-induced neuromuscular block.

Study Design: Prospective, randomized, experimental study.

Animals: A total of six intact healthy adult male Beagle dogs, weighing 9.

View Article and Find Full Text PDF

Accuracy of acceleromyography (AMG) is not be comparable to that of mechanomyography or electromyography (EMG). In particular, the prone position may reduce the accuracy and feasibility of AMG. We developed a new device based on wrist brace to allow free thumb movement and fix the other parts of the hand and wrist.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!