Objective: Balloon-expandable stent valves require flow reduction during implantation (rapid pacing). The present study was designed to compare a self-expanding stent valve with annular fixation versus a balloon-expandable stent valve.

Methods: Implantation of a new self-expanding stent valve with annular fixation (Symetis, Lausanne, Switzerland) was assessed versus balloon-expandable stent valve, in a modified Dynatek Dalta pulse duplicator (sealed port access to the ventricle for transapical route simulation), interfaced with a computer for digital readout, carrying a 25 mm porcine aortic valve. The cardiovascular simulator was programmed to mimic an elderly woman with aortic stenosis: 120/85 mmHg aortic pressure, 60 strokes/min (66.5 ml), 35% systole (2.8 l/min).

Results: A total of 450 cardiac cycles was analysed. Stepwise expansion of the self-expanding stent valve with annular fixation (balloon-expandable stent valve) resulted in systolic ventricular increase from 120 to 121 mmHg (126 to 830+/-76 mmHg)*, and left ventricular outflow obstruction with mean transvalvular gradient of 11+/-1.5 mmHg (366+/-202 mmHg)*, systolic aortic pressure dropped distal to the valve from 121 to 64.5+/-2 mmHg (123 to 55+/-30 mmHg) N.S., and output collapsed to 1.9+/-0.06 l/min (0.71+/-0.37 l/min* (before complete obstruction)). No valve migration occurred in either group. (*=p<0.05).

Conclusions: Implantation of this new self-expanding stent valve with annular fixation has little impact on haemodynamics and has the potential for working heart implantation in vivo. Flow reduction (rapid pacing) is not necessary.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejcts.2009.01.052DOI Listing

Publication Analysis

Top Keywords

stent valve
24
valve annular
16
annular fixation
16
balloon-expandable stent
16
self-expanding stent
12
valve
9
stent
8
versus balloon-expandable
8
aortic pressure
8
mmhg
5

Similar Publications

Background And Objectives: Currently, there remains a paucity of research on the deformation and valve function of transcatheter heart valves (THV) in patients with aortic regurgitation (AR) following transcatheter aortic valve replacement (TAVR). This study aimed to thoroughly explore the correlation between THV deformation and postoperative hemodynamics in these patients.

Methods: In this study, we assessed 39 AR patients treated with the J-Valve THV system during TAVR.

View Article and Find Full Text PDF

The use of the snare catheter (SC) technique has been described in the field of interventional cardiology, in particular in the retrieval of a lost device, for example, a dislodged coronary stent, broken coronary wire, and so forth. In the transcatheter aortic valve replacement (TAVR) procedure, some cases have been observed where the anatomy is challenging or there are scenarios where some complications occur during the procedure, which make it necessary to use some tools to achieve the success of the procedure. The SC has shown are very useful either to achieve the ascent of the valve to the annular plane in complex anatomies or as a rescue measure in the event of complications that may arise after valve implantation.

View Article and Find Full Text PDF

Valve Frame Geometry and Arrhythmia Risk Following Self-Expanding Transcatheter Pulmonary Valve Replacement.

Pediatr Cardiol

January 2025

School of Medicine, Children's Hospital Colorado, The Heart Institute, University of Colorado Denver|Anschutz Medical Campus, 13123 E 16Th Ave, Box 100, Aurora, CO, 80045-2560, USA.

Though major complications during transcutaneous pulmonary valve replacement (TPVR) are rare, clinically-significant ventricular arrythmia (CSVA) has been reported following self-expanding valve placement. We assess whether alterations in valve frame dimensions and geometry within the right ventricular outflow tract (RVOT) post-implantation in patients who underwent TPVR with Harmony TPV25 or Alterra contribute to CSVA risk. A single center review was performed of patients who underwent TPVR with either Harmony TPV25 or Alterra Pre-stent between August 2019 and April 2023.

View Article and Find Full Text PDF

Introduction: Coarctation of the aorta is a congenital narrowing of the thoracic aorta associated with hypertension and significant pressure gradients across the coarctation site. Coarctoplasty by percutaneous approach is the preferred method of treatment. However, complications like stent dislodgement may result and must be immediately managed to prevent adverse outcomes.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!