The metal-backed patella. An invitation for failure?

J Arthroplasty

Department of Orthopaedics, Fitzsimons Army Medical Center, Aurora, Colorado.

Published: September 1991

Thirty-one cases of failed metal-backed patellar components requiring revision surgery are reviewed. Two modes of implant failure onset were noted. Within each group, delays in patient presentation and diagnosis of the failed component were encountered. The syndrome of the failed metal-backed patellar component is proposed. The authors conclude that: (1) more of these failures will be seen in the future; (2) this patient population needs appropriate counsel concerning the signs and symptoms of this syndrome as well as more frequent follow-up evaluations; (3) surgeons treating this complication should be prepared to revise all components; and (4) the use of metal-backed patellar components should be discontinued.

Download full-text PDF

Source
http://dx.doi.org/10.1016/s0883-5403(06)80168-3DOI Listing

Publication Analysis

Top Keywords

metal-backed patellar
12
failed metal-backed
8
patellar components
8
metal-backed
4
metal-backed patella
4
patella invitation
4
invitation failure?
4
failure? thirty-one
4
thirty-one cases
4
cases failed
4

Similar Publications

Early Shear Failure of a 3-Peg Modified Dome Patellar Implant.

Arthroplast Today

August 2024

Department of Orthopaedic Surgery, UCLA Health, Santa Monica, CA, USA.

Patellar failure in total knee arthroplasty is a major source of complications postoperatively. Previous patellar failure reports commonly cited dissociations of modular and metal-backed patellar implants. However, mechanical breakage of monoblock all-polyethylene patellar implants is very rare.

View Article and Find Full Text PDF

Metallosis is a known yet rare late complication of unicompartmental and total knee arthroplasty (TKA), usually secondary to either metal-backed patellar component failure, mobile-bearing polyethylene dislocation, or catastrophic polyethylene failure and wear through. The majority of literature surrounding metallosis has been published in relation to total hip arthroplasty (THA) metal on metal bearing wear or mechanically assisted crevice corrosion.This case report describes the development of metallosis in a 77-year-old male patient with advanced (Kellgren-Lawrence Grade 4) osteoarthritis with associated valgus deformity, who underwent index TKA with a semiconstrained revision knee system due to intraoperative medial collateral ligament laxity.

View Article and Find Full Text PDF

Aims: The use of cementless total knee arthroplasty (TKA) components has increased during the past decade. The initial design of cementless metal-backed patellar components had shown high failure rates due to many factors. The aim of this study was to evaluate the clinical results of a second-generation cementless, metal-backed patellar component of a modern design.

View Article and Find Full Text PDF

Background: Cementless total knee arthroplasty (TKA) has become increasingly popular. Some surgeons are concerned about pain, implant stability, and metal-backed patellar component survivorship. This study investigated the outcomes of cementless compared with cemented TKA in bilateral cases.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!