To eliminate the diastereomer interference on Telcagepant (MK-0974) determination during clinical study support, on-line high turbulent-flow liquid chromatography (HTLC) methods, HTLC-A and HTLC-B that covered dynamic range of 0.5-500 nM and 5-5000 nM, respectively, were developed. To meet the requirement of rapid assay transfer among multiple laboratories and analysts, a solid-phase extraction (SPE) assay was derived from the existing HTLC-B assay under the same dynamic range. The on-line HTLC assays were achieved through direct injection of plasma samples, extraction of analyte with a Cohesive C18 column (50 mm x 0.5 mm, 50 microm), followed by HPLC separation on a FluoPhase RP column (100 mm x 2.1 mm, 5 microm) and MS/MS detection. The off-line SPE assay used Waters Oasis HLB microElution plate to extract the analytes from plasma matrix before injecting on a FluoPhase RP column (150 mm x 2.1 mm, 5 microm) for LC-MS/MS analysis. Under both on-line and off-line assay conditions, the diastereomer 1c was chromatographically separated from MK-0974. Cross-validation with the pooled samples demonstrated that both on-line and off-line assays provided comparable data with a difference of < 2.6%. The assays were proved to be specific, accurate and reliable, and have been used to support multiple clinical studies. The pros and cons of on-line and off-line assays with regard to man power involved in sample preparation, total analysis time, carryover, cost efficiency, and the requirement for assay transfer are discussed.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jchromb.2009.04.006DOI Listing

Publication Analysis

Top Keywords

on-line off-line
12
diastereomer interference
8
telcagepant mk-0974
8
solid-phase extraction
8
dynamic range
8
assay transfer
8
spe assay
8
fluophase column
8
off-line assays
8
on-line
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!