Purpose: To quantitatively assess and compare the anterior chamber volume (ACV) by different anterior segment imaging techniques.

Methods: Forty healthy eyes of 40 patients were recruited and three consecutive measurements of ACV were determined prospectively utilizing SL-OCT and Pentacam. Statistical significance of interdevice differences between measurements was evaluated by unpaired t test and Bland-Altman analysis. The repeatability of three consecutive measurements was analyzed by repeated measured analysis of variance.

Results: Eighty eyes of 40 patients (18 female, 22 male) with a mean spherical error of -0.55+/-1.4 diopters were included in the study. Mean ACV were detected as 163.34+/-29.76 mm3 by Pentacam and 164.14+/-28.44 mm3 by SL-OCT (unpaired t test, p=0.872). Pentacam and SL-OCT displayed a high intrasession repeatability (repeated measured analysis of variance, p>0.05). ACV values detected by SL-OCT were clinically interchangeable with Pentacam.

Conclusions: Noncontact measurements of ACV using SL-OCT and Pentacam were easy to handle and demonstrated good repeatability. The tested devices were regarded as compatible in measuring ACV.

Download full-text PDF

Source
http://dx.doi.org/10.1177/112067210901900314DOI Listing

Publication Analysis

Top Keywords

anterior chamber
8
chamber volume
8
eyes patients
8
three consecutive
8
consecutive measurements
8
measurements acv
8
sl-oct pentacam
8
unpaired test
8
repeated measured
8
measured analysis
8

Similar Publications

Objective: This study investigates the refractive accuracy of eight intraocular lens (IOL) power calculation formulas in patients with postoperative refractive surprise after phacoemulsification. It aims to determine if a different formula could result in better refractive outcomes in these eyes.

Methods And Analysis: We retrospectively reviewed consecutive patients undergoing uncomplicated phacoemulsification as a sole procedure between March 2007 and September 2020 at the University of Washington by glaucoma subspecialists as part of a study investigating cataract surgery in normal eyes.

View Article and Find Full Text PDF

Nd:YAG Laser Application for the Treatment of Retained Lens Fragment in the Anterior Chamber Following Cataract Surgery.

Turk J Ophthalmol

December 2024

University of Health Sciences Türkiye, Ulucanlar Eye Training and Research Hospital, Clinic of Ophthalmology, Ankara, Türkiye.

Cataract surgery is the most frequently performed surgery worldwide. Although it is an effective surgical treatment option for improving patients' visual acuity, various complications can occur postoperatively. One such complication is the presence of retained lens material in the anterior chamber, which can lead to intraocular inflammation, increased intraocular pressure, corneal edema, and endothelial cell loss.

View Article and Find Full Text PDF

To observe the characteristics of ocular biological parameters in children with transfusion-dependent β-thalassemia (TDT) and the effect of iron chelator treatment on them. This was a cross-sectional study. Thirty-two children with TDT (TDT group) and 64 healthy children (control group) who were treated in the Affiliated Hospital of Guangdong Medical University from October 2022 to June 2023 were included.

View Article and Find Full Text PDF

Comparison of the Efficacy of Endoscopic Continuous Perfusion Combined With Y-Shaped Incision and Microscopic Retroauricular Incision in the Treatment of Attic Cholesteatoma: A Randomized Prospective Study.

Ear Nose Throat J

January 2025

Department of Otolaryngology, Northern Jiangsu People's Hospital, Institute of Translational Medicine, Medical College, Yangzhou University, Yangzhou, PR China.

This study aimed to compare the efficacy of continuous perfusion of underwater bone grinding combined with a -shaped incision versus a microscopic posterior ear incision in the treatment of attic cholesteatoma. Clinical trials were prospective studies from the Northern Jiangsu People's Hospital. Adult patients with middle ear cholesteatoma requiring ear surgery agreed to participate between September 2019 and September 2023 (age > 18).

View Article and Find Full Text PDF

Purpose: This study evaluates the discrepancies between ACV measurements obtained from the Heidelberg Anterion and Zeiss IOLMaster 700 and investigates the significance of ACV and other ocular biometry parameters.

Patients And Methods: To investigate intraocular fluid circulation, a robust formula was developed for ACV measurement using the Zeiss IOLMaster 700. A pilot study was conducted to validate this formula, which relied on WTW, CCT, and ACD.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!