Design and implementation of a compact low-dose diffraction enhanced medical imaging system.

Acad Radiol

Departments of Radiology and Biomedical Engineering, School of Medicine, University of North Carolina at Chapel Hill, Chapel Hill, NC 27599-7000, USA.

Published: August 2009

Rationale And Objectives: Diffraction-enhanced imaging (DEI) is a new x-ray imaging modality that differs from conventional radiography in its use of three physical mechanisms to generate contrast. DEI is able to generate contrast from x-ray absorption, refraction, and ultra-small-angle scatter rejection (extinction) to produce high-contrast images with a much lower radiation dose compared to conventional radiography.

Materials And Methods: A prototype DEI system was constructed using a 1-kW tungsten x-ray tube and a single silicon monochromator and analyzer crystal. The monochromator crystal was aligned to reflect the combined Kalpha1 (59.32 keV) and Kalpha2 (57.98 keV) characteristic emission lines of tungsten using a tube voltage of 160 kV. System performance and demonstration of contrast were evaluated using a nylon monofilament refraction phantom, full-thickness breast specimens, a human thumb, and a live mouse.

Results: Images acquired using this system successfully demonstrated all three DEI contrast mechanisms. Flux measurements acquired using this 1-kW prototype system demonstrated that this design can be scaled to use a more powerful 60-kW x-ray tube to generate similar images with an image time of approximately 30 seconds. This single-crystal pair design can be further modified to allow for an array of crystals to reduce clinical image times to <3 seconds.

Conclusions: This paper describes the design, construction, and performance of a new DEI system using a commercially available tungsten anode x-ray tube and includes the first high-quality low-dose diffraction-enhanced images of full-thickness human tissue specimens.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.acra.2009.02.007DOI Listing

Publication Analysis

Top Keywords

generate contrast
8
x-ray tube
8
system demonstrated
8
system
5
design implementation
4
implementation compact
4
compact low-dose
4
low-dose diffraction
4
diffraction enhanced
4
enhanced medical
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!