Rationale And Objectives: To determine whether angled gantry acquisition might be used to image the heart with a shorter scan length and reduced breast exposure during coronary computed tomography angiography.

Materials And Methods: One hundred consecutive coronary computed tomography angiography examinations of female patients were retrospectively evaluated to define the angle between the long axis of the left heart and the axial imaging plane. The scan length required to image the entire left ventricle along with the coronary arteries was measured for an axial scan plane as well as for a scan plane parallel to the long axis of the left heart. The overlap between these imaging volumes and the lower portion of the breast was measured.

Results: The long axis of the left heart varied from 7 degrees to 54 degrees off the axial plane (mean 32 degrees +/- 7 degrees ). The required scan length to include the entire left ventricle and coronary arteries ranged from 8.2 to 12.4 cm (mean, 10.0 +/- 0.9 cm) for the axial scan plane and 5.6-10.1 cm (mean, 7.5 +/- 0.8 cm) for a scan plane parallel to the long axis of the heart (P < .001). cCTA in the axial plane required a 7.4 +/- 1.6 cm overlap with the lower breast, whereas cCTA in the long axis of the heart reduced the overlap to 4.5 +/- 1.8 cm (P < .001).

Conclusions: Using an angled gantry approach, the coronary arteries can be fully imaged in a plane along the long axis of the left heart with a single 10-cm acquisition and with substantial reduction in amount of breast tissue within the irradiated field.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.acra.2009.02.009DOI Listing

Publication Analysis

Top Keywords

long axis
24
axis left
16
left heart
16
scan plane
16
scan length
12
coronary arteries
12
exposure coronary
8
angled gantry
8
coronary computed
8
computed tomography
8

Similar Publications

Background: Closed head injury (CHI) provokes a prominent neuroinflammation that may lead to long-term health consequences. Microglia plays pivotal and complex roles in neuroinflammation-mediated neuronal insult and repair following CHI. We previously reported that induced neural stem cells (iNSCs) can block the effects of CXCL12/CXCR4 signaling on NF-κB activation in activated microglia by CXCR4 overexpression.

View Article and Find Full Text PDF

Cr(VI) is widely used in industry and has high toxicity, making it one of the most common environmental pollutants. Long-term exposure to Cr(VI) can cause metabolic disorders and tissue damage. However, the effects of Cr(VI) on liver and gut microbes in fish have rarely been reported.

View Article and Find Full Text PDF

Is a vertical fracture fragment after indirect reduction acceptable in minimally invasive plate osteosynthesis for acute mid-shaft clavicular fractures?

Injury

January 2025

Department of Orthopedic Surgery, Korea University Anam Hospital, Korea University College of Medicine, 73, Goryeodae-ro, Seongbuk-gu, Seoul 02841, Republic of Korea. Electronic address:

Purpose: Reduction and intraoperative maintenance of fracture fragments during minimally invasive plate osteosynthesis (MIPO) pose technical difficulties, particularly when the interposed fragment is angulated, prompting surgeons to attempt reduction due to concerns about nonunion or malunion. We aimed to compare the clinical and radiological outcomes of MIPO for mid-shaft clavicular fractures based on the reduced status of the interposed fragments.

Method: Fifty-seven patients who underwent MIPO for acute mid-shaft Robinson type 2B clavicular fractures were divided into two groups based on the alignment of the interposed fracture fragment.

View Article and Find Full Text PDF

Film-coupled plasmonic resonators offer efficient platforms for light enhancement due to the excitation of gap surface plasmons (GSPs) at metal-insulator-metal interfaces, where electromagnetic energy is stored within the spacer. In applications like biosensing and spontaneous emission control, spatial overlap between the target molecule and plasmonic hotspots is essential. Here, we propose utilizing the controllable, efficient light enhancement capabilities of a specifically designed GSP disk resonator for biosensing and spontaneous emission enhancement.

View Article and Find Full Text PDF

Scintillation restricts the development of free space optical communication in near-space hypersonic vehicles. We derived analytical formulae for the on-axis scintillation index of a partially coherent flat-topped beam in a bidirectional turbulent atmosphere and plasma link based on the Collins formula. The numerical results indicate that over long-distance propagation ranges, the on-axis scintillation index of a partially coherent flat-topped beam in bidirectional transmission is smaller than that of a Gaussian Schell model beam under the same conditions.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!