Background/objective: Abnormal cerebral haemodynamics in very preterm infants undergoing neonatal intensive care have been associated with adverse outcome, but cerebral blood flow (CBF) is difficult to assess at the cotside. Spatially resolved spectroscopy (SRS) continuously measures cerebral tissue oxygen saturation expressed as tissue oxygenation index (TOI, %), and TOI would reflect changes in CBF with constant cerebral metabolic demand. We aimed to evaluate the relationship between simultaneous measurements of CBF (Transonic System) and TOI (Hamamatsu NIRO-200) in newborn lambs (n = 8). We hypothesised that alterations in CBF of different magnitudes and frequencies would be reflected as changes of TOI in both time domain and frequency domain analyses.

Methods: A silicon cuff positioned around the common brachiocephalic artery was inflated to induce 10-30% reductions in cerebral perfusion pressure so as to produce mild-to-moderate variations of CBF. Relationships between changes in CBF and TOI were evaluated in the time domain (Pearson correlation) and frequency domain (Coherence).

Results: In time domain analysis, there was significant correlation between DeltaTOI (%) and DeltaCBF (%) (R2 = 0.69, P < 0.001). In frequency domain analysis, CBF-TOI coherence was > or = 0.5 at frequencies below 0.1 Hz, but <0.5 at higher frequencies.

Conclusions: We conclude that overall TOI changes are concordant with CBF variations, when arterial oxygen saturation and cerebral oxygen consumption are constant. While TOI reflects CBF, it is more sensitive to variations of CBF of low frequency (<0.1 Hz) than to more rapid, higher frequency changes.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00134-009-1486-4DOI Listing

Publication Analysis

Top Keywords

time domain
12
frequency domain
12
tissue oxygenation
8
spatially resolved
8
resolved spectroscopy
8
cerebral blood
8
blood flow
8
newborn lambs
8
changes cbf
8
domain analysis
8

Similar Publications

Prcis: Cognitive impairment in multiple domains was observed in primary open angle glaucoma patients as compared to age and gender matched healthy controls.

Objective: Evaluation of cognitive impairment in individuals with Primary Open Angle Glaucoma (POAG).

Methods: In this case-control study, individuals with POAG (cases, n=70) were compared with age- and sex-matched healthy individuals (controls, n=70) using detailed ophthalmological evaluation, cognitive assessment and serum cortisol level.

View Article and Find Full Text PDF

The 2024 Zurich perfluorinated compounds (PFCs) summit reiterated the urgent need for non-selective analytical approaches for PFC detection. 19F NMR holds great potential, however, sensitivity limitations lead to long analysis times and/or the possibility of not detecting low concentration species. Steady State Free Precession (SSFP) NMR collects the signal in a steady state regime, allowing 100's of acquisitions in the timespan of a single traditional NMR scan.

View Article and Find Full Text PDF

Background: Changes in cardiac function and structure as well as their association with the cardiac autonomic nervous system remain incompletely characterized in children with stage 5 chronic kidney disease (CKD) receiving hemodialysis (HD).

Methods: A prospective observational cohort study was conducted on 40 Egyptian children with CKD on regular HD compared to 40 age- and sex-matched healthy children. All participants underwent thorough clinical examination, laboratory investigations, 24-h Holter monitoring, and 2D/4D echocardiographic study (conventional and advanced modalities).

View Article and Find Full Text PDF

A solution method for active suppression of reflections in anechoic chambers.

J Acoust Soc Am

January 2025

University of Twente, Faculty of Engineering Technology, Applied Mechanics and Data Analysis, Drienerlolaan 5, 7522 NG Enschede, The Netherlands.

A solution method to improve an anechoic chamber at low frequencies with the use of active noise control is presented. The approach uses the Kirchhoff-Helmholtz integral to compute the reflected sound field resulting from the primary sources together with an algorithm to compute the filter coefficients of a controller driving secondary sources on the walls of the enclosure using reference signals as inputs, which are measured on a contour enclosing the primary sources. A causal frequency domain method with conjugate gradient iterations is derived to determine the controller.

View Article and Find Full Text PDF

Background: Research on the contextual drivers of antimicrobial stewardship (AMS) programme interventions in neonatal units is limited.

Methods: As part of a prospective mixed-methods multidisciplinary neonatal AMS (NeoAMS) interventional study in 14 South African hospitals, we applied a three-phased process to assess implementation barriers and contextual drivers experienced by participating health professionals. The study included: (Phase one; P1) a survey of pre-intervention barriers and enablers; (P2) written feedback during the study intervention phase; and (P3) semi-structured exit interviews.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!