Background: Sirolimus-eluting stents (SES) have incomplete neointimal coverage at 6-month follow up as determined with optical coherence tomography (OCT). The long-term detailed changes of neointima in SES remains to be clarified.

Methods And Results: Serial changes in neointimal coverage of SES from 6 months to 12 months using OCT were examined. Of 21 SES in 13 patients, OCT was used to visualize 2,321 stent struts at 6 months and 2,285 stent struts at 12 months. The frequency of struts without neointimal coverage decreased from 6 months to 12 months (from 10.4 to 5.7%). The frequency of malapposed struts decreased from 6 months to 12 months (from 1.7 to 0.2%). The average thickness of the neointima increased (from 112 +/-123 to 120 +/-130 microm). The frequency of struts located at the side branch orifice without neointima decreased (from 4 out of 17 (24%) to 0 out of 17 (0%)). Complete coverage with neointima was observed in 14% (3 of 21 SES) at 6 months, and 24% (5 of 21 SES) at 12 months.

Conclusions: Additional neointimal coverage was observed between 6 and 12 months, with a small increase in the neointimal thickness. The incidence of complete coverage, however, was still low at 12 months. These findings suggest delayed neointimalization on SES.

Download full-text PDF

Source
http://dx.doi.org/10.1253/circj.cj-08-0746DOI Listing

Publication Analysis

Top Keywords

neointimal coverage
16
months months
12
months
11
delayed neointimalization
8
sirolimus-eluting stents
8
optical coherence
8
coherence tomography
8
ses months
8
stent struts
8
struts months
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!