An easy and fast Fourier transform continuous cyclic voltammetric technique (FFTCV) for monitoring of ultra trace amounts of cyclizine in a flow-injection system has been introduced in this work. The potential waveform, which was applied continuously on an Au disk microelectrode (12.5 microm in radius) consisted of the potential steps for cleaning, accumulation and potential ramp. The proposed detection method has some advantages, the greatest of which are as follows: first, it is no longer necessary to remove oxygen from the analyte solution and second, this is a very fast and appropriate technique for determination of the drug compound in a wide variety of chromatographic analysis methods. The detection limit for cyclizine was 1.8 ng ml(-1). The relative standard deviation (RSD) of the proposed technique at 5.0 x 10(-7) was 2.0 for 10 runs. The influences of pH of eluent, accumulation potential, sweep rate, and accumulation time on the determination of the cyclizine were considered. The proposed method was applied to the determination of cyclizine in a pharmaceutical preparation.

Download full-text PDF

Source
http://dx.doi.org/10.2116/analsci.25.505DOI Listing

Publication Analysis

Top Keywords

fast fourier
8
fourier transform
8
transform continuous
8
continuous cyclic
8
accumulation potential
8
determination cyclizine
8
cyclizine
5
application adsorptive
4
adsorptive voltammetry
4
voltammetry detection
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!